Your browser doesn't support javascript.
loading
A new complimentary web-based tool for manual analysis of microcirculation videos: Validation of the Capillary Mapper against the current gold standard AVA 3.2.
Hessler, Michael; Arnemann, Philip-Helge; Zamit, Frédéric; Seidel, Laura; Kampmeier, Tim-Gerald; Kathöfer, Ulrich; Morelli, Andrea; Ince, Can; Rehberg, Sebastian W; Ertmer, Christian.
Afiliação
  • Hessler M; Department of Anesthesiology, Intensive Care and Pain Therapy, University of Muenster Medical Centre, Muenster, Germany.
  • Arnemann PH; Department of Anesthesiology, Intensive Care and Pain Therapy, University of Muenster Medical Centre, Muenster, Germany.
  • Zamit F; IT-Center Research and Education, Medical Faculty of Muenster, University of Muenster, Muenster, Germany.
  • Seidel L; Department of Anesthesiology, Intensive Care and Pain Therapy, University of Muenster Medical Centre, Muenster, Germany.
  • Kampmeier TG; Department of Anesthesiology, Intensive Care and Pain Therapy, University of Muenster Medical Centre, Muenster, Germany.
  • Kathöfer U; IT-Center Research and Education, Medical Faculty of Muenster, University of Muenster, Muenster, Germany.
  • Morelli A; Department of Anesthesiology and Intensive Care, University of Rome "La Sapienza", Rome, Italy.
  • Ince C; Department of Intensive Care, Erasmus MC, University Medical Center, Rotterdam, The Netherlands.
  • Rehberg SW; Department of Anesthesiology, Intensive Care, Emergency Medicine, Transfusion Medicine and Pain Therapy, Protestant Hospital of the Bethel Foundation, Bielefeld, Germany.
  • Ertmer C; Department of Anesthesiology, Intensive Care and Pain Therapy, University of Muenster Medical Centre, Muenster, Germany.
Microcirculation ; 25(8): e12505, 2018 11.
Article em En | MEDLINE | ID: mdl-30192423
ABSTRACT

OBJECTIVE:

The aim of the current study was to compare a newly developed web-based freely accessible software program for manual analysis of the microcirculation, the Capillary Mapper (CM), with AVA 3.2 software (AVA; MicroVision Medical B.V., Amsterdam, The Netherlands), which is the current gold standard for analysis of microcirculation videos.

METHODS:

A web-based software program was developed, which enables manual analysis of videos of the microcirculation to be carried out according to recommendations of the 2018 consensus conference. A set of 50 high quality microcirculation videos was analyzed with AVA and CM with respect to total vessel density, perfused vessel density, proportion of perfused vessels, and the microvascular flow index.

RESULTS:

Comparison of the mean values derived from manual analysis with CM and AVA revealed no significant differences in microcirculatory variables. Analysis according to Bland and Altman revealed an acceptable bias between manual analysis with the CM and AVA for all variables tested with sufficient limits of agreement. The analysis of intraclass correlation showed "excellent" agreement for all microcirculatory variables analyzed.

CONCLUSIONS:

The newly developed CM was successfully validated for manual analyses of microcirculation videos against the current gold standard, the software AVA 3.2.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Gravação de Videoteipe / Microcirculação Limite: Humans Idioma: En Ano de publicação: 2018 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Gravação de Videoteipe / Microcirculação Limite: Humans Idioma: En Ano de publicação: 2018 Tipo de documento: Article