Evaluating clinical observation versus Spatial Frequency Domain Imaging (SFDI), Laser Speckle Imaging (LSI) and thermal imaging for the assessment of burn depth.
Ponticorvo, Adrien; Rowland, Rebecca; Baldado, Melissa; Burmeister, David M; Christy, Robert J; Bernal, Nicole P; Durkin, Anthony J.
Afiliação
  • Ponticorvo A; Beckman Laser Institute and Medical Clinic, University of California, 1002 Health Sciences Road East, Irvine, CA 92617, United States.
  • Rowland R; Beckman Laser Institute and Medical Clinic, University of California, 1002 Health Sciences Road East, Irvine, CA 92617, United States.
  • Baldado M; Beckman Laser Institute and Medical Clinic, University of California, 1002 Health Sciences Road East, Irvine, CA 92617, United States.
  • Burmeister DM; United States Army Institute of Surgical Research, 3650 Chambers Pass, Fort Sam Houston, TX, 78234, United States.
  • Christy RJ; United States Army Institute of Surgical Research, 3650 Chambers Pass, Fort Sam Houston, TX, 78234, United States.
  • Bernal NP; UC Irvine Regional Burn Center, Department of Surgery, 333 City Boulevard West, Suite 705, Orange, CA 92868, United States.
  • Durkin AJ; Beckman Laser Institute and Medical Clinic, University of California, 1002 Health Sciences Road East, Irvine, CA 92617, United States; Department of Biomedical Engineering, University of California, 3120 Natural Sciences II, Irvine, CA 92697, United States. Electronic address: adurkin@uci.edu.
Burns ; 45(2): 450-460, 2019 03.
Article em En | MEDLINE | ID: mdl-30327232
ABSTRACT
While clinical examination is needed for burn severity diagnosis, several emerging technologies aim to quantify this process for added objectivity. Accurate assessments become easier after burn progression, but earlier assessments of partial thickness burn depth could lead to earlier excision and grafting and subsequent improved healing times, reduced rates of scarring/infection, and shorter hospital stays. Spatial Frequency Domain Imaging (SFDI), Laser Speckle Imaging (LSI) and thermal imaging are three non-invasive imaging modalities that have some diagnostic ability for noninvasive assessment of burn severity, but have not been compared in a controlled experiment. Here we tested the ability of these imaging techniques to assess the severity of histologically confirmed graded burns in a swine model. Controlled, graded burn wounds, 3cm in diameter were created on the dorsum of Yorkshire pigs (n=3, 45-55kg) using a custom-made burn tool that ensures consistent pressure has been employed by various burn research groups. For each pig, a total of 16 burn wounds were created on the dorsal side. Biopsies were taken for histological analysis to verify the severity of the burn. Clinical analysis, SFDI, LSI and thermal imaging were performed at 24 and 72h after burn to assess the accuracy of each imaging technique. In terms of diagnostic accuracy, using histology as a reference, SFDI (85%) and clinical analysis (83%) performed significantly better that LSI (75%) and thermography (73%) 24h after the burn. There was no statistically significant improvement from 24 to 72h across the different imaging modalities. These data indicate that these imaging modalities, and specifically SFDI, can be added to the burn clinicians' toolbox to aid in early assessment of burn severity.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Queimaduras Tipo de estudo: Prognostic_studies Limite: Animals Idioma: En Ano de publicação: 2019 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Queimaduras Tipo de estudo: Prognostic_studies Limite: Animals Idioma: En Ano de publicação: 2019 Tipo de documento: Article