Your browser doesn't support javascript.
loading
[Surface microstructure and cyclic fatigue resistance of electro discharged machining nickel-titanium endodontic instrument].
Wang, F F; Yang, Y J; Hou, X M.
Afiliação
  • Wang FF; Second Clinical Division, Peking University School and Hospital of Stomatology & National Clinical Research Center for Oral Diseases & National Engineering Laboratory for Digital and Material Technology of Stomatology & Beijing Key Laboratory of Digital Stomatology, Beijing 100101, China; Department of Stomatology, Xuanwu Hospital, Capital Medical University, Beijing 100053, China.
  • Yang YJ; Department of Stomatology, Beijing Xicheng District Xinjiekou Community Health Service Center, Beijing 100035, China.
  • Hou XM; Second Clinical Division, Peking University School and Hospital of Stomatology & National Clinical Research Center for Oral Diseases & National Engineering Laboratory for Digital and Material Technology of Stomatology & Beijing Key Laboratory of Digital Stomatology, Beijing 100101, China.
Beijing Da Xue Xue Bao Yi Xue Ban ; 50(5): 876-881, 2018 Oct 18.
Article em Zh | MEDLINE | ID: mdl-30337751
ABSTRACT

OBJECTIVE:

To compare the surface microstructures and cyclic fatigue resistance of HyFlex EDM with HyFlex CM and HyFlex NT.

METHODS:

Twelve 25 mm-long 25#/- HyFlex EDM and twelve 25 mm-long 25#/0.06 taper HyFlex CM or HyFlex NT were selected. The surface microstructure of the 2 instruments which were randomly selected from each group was observed by using scanning electron microscope (SEM). The remaining 10 instruments from each group were submitted to the cyclic fatigue test by using a simulated stainless steel root canal with 60° angle of curvature and curvature radius of 3.5 mm. The time till fracture was recorded, the number of cyclic fatigue (NCF) was calculated, the length of fracture fragment was evaluated, and the topographic features were analyzed by using SEM. The data were analyzed by using one-way analysis via SPSS 23.0 software. The statistical significance level was set at 0.05.

RESULTS:

SEM observation of the surface and microstructural characterization revealed peculiar melting appearance and evenly distributed micropores on the surface of HyFlex EDM while observations of the surface of both HyFlex CM and HyFlex NT demonstrated machining grooves and irregularities. The NCF of HyFlex EDM was 838±223, which was significantly higher than not only that of HyFlex CM (582±99), but also that of HyFlex NT (81±20) (P<0.05), and the difference between the latter two groups was significant, as well (P<0.05). The length of fractured fragments of HyFlex EDM was (7.27±0.28) mm, which was significantly longer than the HyFlex CM and HyFlex NT, with values of (6.72±0.26) mm and (6.62±0.37) mm, respectively (P<0.05). Topographic features demonstrated typical cyclic fatigue for all the three groups while the region of crack origins was more for HyFlex EDM and HyFlex CM than that of HyFlex NT. As far as the dimple area was concerned, that of HyFlex EDM was deeper and larger than that of HyFlex CM and HyFlex NT.

CONCLUSION:

Within the limitations of this study, electro-discharge machining leads to peculiar melting appearance with micropores instead of machining grooves and irregularities on the surface of HyFlex EDM, which may be the reason why HyFlex EDM exhibits significantly better cyclicfatigue resistance than HyFlex CM and HyFlex NT.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Ligas Dentárias / Instrumentos Odontológicos / Falha de Equipamento Limite: Humans Idioma: Zh Ano de publicação: 2018 Tipo de documento: Article
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Ligas Dentárias / Instrumentos Odontológicos / Falha de Equipamento Limite: Humans Idioma: Zh Ano de publicação: 2018 Tipo de documento: Article