A Calibration Device to Compare Body Plethysmographs Among Pediatric Lung Function Laboratories.
Demoulin, Bruno; Ioan, Iulia; Duvivier, Claude; Bonabel, Claude; Schweitzer, Cyril; Marchal, François; Demoulin-Alexikova, Silvia.
Afiliação
  • Demoulin B; EA 3450 DevAH, Department of Physiology, Faculty of Medicine, University of Lorraine, Nancy, France.
  • Ioan I; EA 3450 DevAH, Department of Physiology, Faculty of Medicine, University of Lorraine, Nancy, France.
  • Duvivier C; Department of Pediatric Respiratory Function Testing, Children's Hospital, Vandoeuvre, France.
  • Bonabel C; EA 3450 DevAH, Department of Physiology, Faculty of Medicine, University of Lorraine, Nancy, France.
  • Schweitzer C; Department of Pediatric Respiratory Function Testing, Children's Hospital, Vandoeuvre, France.
  • Marchal F; EA 3450 DevAH, Department of Physiology, Faculty of Medicine, University of Lorraine, Nancy, France.
  • Demoulin-Alexikova S; Department of Pediatric Respiratory Function Testing, Children's Hospital, Vandoeuvre, France.
Front Physiol ; 9: 1408, 2018.
Article em En | MEDLINE | ID: mdl-30356753
Multi-center studies in specific airway resistance have shown significant inter laboratory variability. Comparison of plethysmographic equipment using a lung model easily transportable from one site to another should be of help to international normative studies. A resistor made of parallel capillary tubes - insuring adequate linearity within 1 L/sec - was connected to a glass bottle. Thermal time constants were measured while the bottle was empty and while stuffed with steel wool. In the latter, isothermal condition was estimated to occur only at very low frequency (around 0.01 Hz) and gas compression was polytropic up to 0.6 Hz. With the empty analog, adiabatic gas compression was estimated to occur at frequencies ≥0.2 Hz, and more accurate volume estimation was obtained. The empty analog volume and specific resistance measured in a body plethysmograph on different days indicated within 5% accuracy as well as intersession repeatability. It is concluded that a physical analog built out of simple material provides accurate measurements of specific resistance. The apparatus should be of help to compare plethysmographic equipments from different laboratories.
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Tipo de estudo: Clinical_trials / Prognostic_studies Idioma: En Ano de publicação: 2018 Tipo de documento: Article