International variation in radiation dose for computed tomography examinations: prospective cohort study.
Smith-Bindman, Rebecca; Wang, Yifei; Chu, Philip; Chung, Robert; Einstein, Andrew J; Balcombe, Jonathan; Cocker, Mary; Das, Marcos; Delman, Bradley N; Flynn, Michael; Gould, Robert; Lee, Ryan K; Nelson, Thomas; Schindera, Sebastian; Seibert, Anthony; Starkey, Jay; Suntharalingam, Saravanabavaan; Wetter, Axel; Wildberger, Joachim E; Miglioretti, Diana L.
Afiliação
  • Smith-Bindman R; Department of Radiology and Biomedical Imaging Epidemiology and Biostatistics, Obstetrics Gynecology and Reproductive Sciences, Philip R Lee Institute for Health Policy Studies, University of California San Francisco, San Francisco, CA, USA Rebecca.Smith-Bindman@ucsf.edu.
  • Wang Y; Department of Radiology and Biomedical Imaging, University of California San Francisco, San Francisco, CA, USA.
  • Chu P; Department of Radiology and Biomedical Imaging, University of California San Francisco, San Francisco, CA, USA.
  • Chung R; Department of Demography, University of California Berkeley, Berkeley, CA, USA.
  • Einstein AJ; Department of Medicine, Division of Cardiology and Department of Radiology, Columbia University Medical Center, New York, NY, USA.
  • Balcombe J; Assuta Medical Center, Tel Aviv-Yafo, Israel.
  • Cocker M; Department of Medical Physics, Churchill Hospital, Oxford University Hospitals NHS Foundation Trust, Oxford, UK.
  • Das M; Department of Diagnostic and Interventional Radiology, Helios Kliniken Duisburg, Duisburg, Germany.
  • Delman BN; Department of Radiology and Nuclear Medicine, Maastricht University Medical Center, Maastricht, Netherlands.
  • Flynn M; Department of Radiology, Icahn School of Medicine at Mount Sinai, New York, NY, USA.
  • Gould R; Henry Ford Health System, Radiology and Public Health Sciences, Detroit, MI, USA.
  • Lee RK; Department of Radiology and Biomedical Imaging, University of California San Francisco, San Francisco, CA, USA.
  • Nelson TR; Department of Radiology, Einstein Healthcare Network, New York, NY, USA.
  • Schindera S; University of California San Diego, San Diego, CA, USA.
  • Seibert A; Cantonal Hospital Aarau, Aarau, Switzerland.
  • Starkey J; Department of Radiology, University California Davis Medical Center, Sacramento, CA, USA.
  • Suntharalingam S; St Luke's International Hospital, Tokyo, Japan.
  • Wetter A; Department of Diagnostic and Interventional Radiology and Neuroradiology, University Hospital Essen, Essen, Germany.
  • Wildberger JE; Institute of Diagnostic and Interventional Radiology and Neuroradiology, Essen University, Essen, Germany.
  • Miglioretti DL; Department of Radiology and Nuclear Medicine, Maastricht University Medical Center, Maastricht, Netherlands.
BMJ ; 364: k4931, 2019 01 02.
Article em En | MEDLINE | ID: mdl-30602590
ABSTRACT

OBJECTIVE:

To determine patient, institution, and machine characteristics that contribute to variation in radiation doses used for computed tomography (CT).

DESIGN:

Prospective cohort study.

SETTING:

Data were assembled and analyzed from the University of California San Francisco CT International Dose Registry.

PARTICIPANTS:

Standardized data from over 2.0 million CT examinations of adults who underwent CT between November 2015 and August 2017 from 151 institutions, across seven countries (Switzerland, Netherlands, Germany, United Kingdom, United States, Israel, and Japan). MAIN OUTCOME

MEASURES:

Mean effective doses and proportions of high dose examinations for abdomen, chest, combined chest and abdomen, and head CT were determined by patient characteristics (sex, age, and size), type of institution (trauma center, care provision 24 hours per day and seven days per week, academic, private), institutional practice volume, machine factors (manufacturer, model), country, and how scanners were used, before and after adjustment for patient characteristics, using hierarchical linear and logistic regression. High dose examinations were defined as CT scans with doses above the 75th percentile defined during a baseline period.

RESULTS:

The mean effective dose and proportion of high dose examinations varied substantially across institutions. The doses varied modestly (10-30%) by type of institution and machine characteristics after adjusting for patient characteristics. By contrast, even after adjusting for patient characteristics, wide variations in radiation doses across countries persisted, with a fourfold range in mean effective dose for abdomen CT examinations (7.0-25.7 mSv) and a 17-fold range in proportion of high dose examinations (4-69%). Similar variation across countries was observed for chest (mean effective dose 1.7-6.4 mSv, proportion of high dose examinations 1-26%) and combined chest and abdomen CT (10.0-37.9 mSv, 2-78%). Doses for head CT varied less (1.4-1.9 mSv, 8-27%). In multivariable models, the dose variation across countries was primarily attributable to institutional decisions regarding technical parameters (that is, how the scanners were used).

CONCLUSIONS:

CT protocols and radiation doses vary greatly across countries and are primarily attributable to local choices regarding technical parameters, rather than patient, institution, or machine characteristics. These findings suggest that the optimization of doses to a consistent standard should be possible. STUDY REGISTRATION Clinicaltrials.gov NCT03000751.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Doses de Radiação / Tomografia Computadorizada por Raios X Tipo de estudo: Clinical_trials / Etiology_studies / Guideline / Observational_studies / Prognostic_studies Limite: Adolescent / Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2019 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Doses de Radiação / Tomografia Computadorizada por Raios X Tipo de estudo: Clinical_trials / Etiology_studies / Guideline / Observational_studies / Prognostic_studies Limite: Adolescent / Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2019 Tipo de documento: Article