A Monte Carlo detector response model for the IRIS PET preclinical scanner.
Pilleri, Alessandro; Camarlinghi, Niccolò; Del Guerra, Alberto; Sportelli, Giancarlo; Belcari, Nicola.
Afiliação
  • Pilleri A; University of Pisa, Department of Physics, Largo Bruno Pontecorvo 3, Pisa 56127, Italy. Electronic address: alessandro.pilleri@df.unipi.it.
  • Camarlinghi N; University of Pisa, Department of Physics, Largo Bruno Pontecorvo 3, Pisa 56127, Italy; INFN Sezione Pisa, Largo Bruno Pontecorvo 3, Pisa 56127, Italy.
  • Del Guerra A; University of Pisa, Department of Physics, Largo Bruno Pontecorvo 3, Pisa 56127, Italy; INFN Sezione Pisa, Largo Bruno Pontecorvo 3, Pisa 56127, Italy.
  • Sportelli G; University of Pisa, Department of Physics, Largo Bruno Pontecorvo 3, Pisa 56127, Italy; INFN Sezione Pisa, Largo Bruno Pontecorvo 3, Pisa 56127, Italy.
  • Belcari N; University of Pisa, Department of Physics, Largo Bruno Pontecorvo 3, Pisa 56127, Italy; INFN Sezione Pisa, Largo Bruno Pontecorvo 3, Pisa 56127, Italy.
Phys Med ; 57: 107-114, 2019 Jan.
Article em En | MEDLINE | ID: mdl-30738514
ABSTRACT
PET preclinical studies require high spatial resolution due to the limited size of the animal under investigation. To achieve this target, iterative image reconstruction algorithms are commonly preferred over the analytical methods because they offer the possibility of accurately modeling the whole imaging process. In this work, we propose an accurate factorized system matrix for the INVISCAN IRIS preclinical PET scanner to be used with an iterative algorithm. The model includes two components the geometric component and the detector response of the system. The main innovative aspect of the work is the creation of the detector matrix using a Monte Carlo simulation, with a particular focus on the optimization of the simulation process to reduce the calculation time. The new system model is compared with the current IRIS model to evaluate the image quality, following the NEMA Standards NU 4-2008. The comparison showed an enhancement of the image quality, in terms of uniformity and recovery coefficients. This work confirms that the inclusion of the detector response into the system model leads to improved reconstruction results.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Método de Monte Carlo / Tomografia por Emissão de Pósitrons combinada à Tomografia Computadorizada / Modelos Teóricos Tipo de estudo: Health_economic_evaluation / Prognostic_studies Limite: Animals Idioma: En Ano de publicação: 2019 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Método de Monte Carlo / Tomografia por Emissão de Pósitrons combinada à Tomografia Computadorizada / Modelos Teóricos Tipo de estudo: Health_economic_evaluation / Prognostic_studies Limite: Animals Idioma: En Ano de publicação: 2019 Tipo de documento: Article