Your browser doesn't support javascript.
loading
Reverse torque evaluation in indexed and nonindexed abutments of Morse Taper implants in a mechanical fatigue test.
Martins, Caio Marques; Ramos, Elimário Venturin; Kreve, Simone; de Carvalho, Geraldo Alberto Pinheiro; Franco, Aline Batista Gonçalves; de Macedo, Luís Guilherme Scavone; Silva, Alecsandro de Moura; Dias, Sérgio Candido.
Afiliação
  • Martins CM; Department of Prosthodontics and Restorative Dentistry, São Leopoldo Mandic Dental Research Center, Campinas, Brazil.
  • Ramos EV; Department of Prosthodontics and Restorative Dentistry, São Leopoldo Mandic Dental Research Center, Campinas, Brazil.
  • Kreve S; Department of Prosthodontics and Restorative Dentistry, São Leopoldo Mandic Dental Research Center, Campinas, Brazil.
  • de Carvalho GAP; Department of Prosthodontics and Restorative Dentistry, São Leopoldo Mandic Dental Research Center, Campinas, Brazil.
  • Franco ABG; Department of Prosthodontics and Restorative Dentistry, São Leopoldo Mandic Dental Research Center, Campinas, Brazil.
  • de Macedo LGS; Department of Oral Rehabilitation, UNESP, São José dos Campos, São Paulo, Brazil.
  • Silva AM; Department of Materials Odontology and Prosthesis, UNESP, São José dos Campos, São Paulo, Brazil.
  • Dias SC; Department of Prosthodontics and Restorative Dentistry, São Leopoldo Mandic Dental Research Center, Campinas, Brazil.
Dent Res J (Isfahan) ; 16(2): 110-116, 2019.
Article em En | MEDLINE | ID: mdl-30820205
ABSTRACT

BACKGROUND:

This experimental study assessed reverse torque of indexed and nonindexed (NI) abutments in Morse Taper (MT) implants in a mechanical fatigue test. MATERIALS AND

METHODS:

In this experimental study It was used 37 implants MT and over them installed Pilar Flex abutments (4.8 mm × 6 mm × 1.5 mm). The groups were as follows Group A used 19 MT implants with a NI Pilar Flex abutment loaded with 32 N/cm and Group B used 18 MT implants with an indexed (IN) Pilar Flex abutment loaded with 20 N/cm. The abutments were tested according to ISO standard 14801/2007. The specimens were installed at 30° from the axial axis and underwent a 133 N load, 4 Hz frequency, and 1,000,000 cycles. Once the test was completed, the reverse torque was provided by an electronic torque meter. Data were submitted to statistical analysis using the t-test for independent samples and paired t-test. The significance level was considered P < 0.05.

RESULTS:

Results obtained showed that the indexed Pilar Flex abutment had a percentage of torque loss from the initial torque of 49% and the NI Pilar Flex abutment lost 14%. Paired Student's t-tests revealed that for both NI (P < 0.001) and indexed (P < 0.001) abutments, the counter torque values were significantly lower than those applied at the initial torque.

CONCLUSION:

According on the methodology used, the NI Pilar Flex abutment was more effective regarding the reverse torque in single-unit implant prostheses versus the indexed Pilar Flex abutment. A greater loosening in the indexed Pilar Flex abutment retaining screw was noted in the reverse torque test, and the Pilar Flex abutment failed to show good outcomes related to the cold welding effect.
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Ano de publicação: 2019 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Ano de publicação: 2019 Tipo de documento: Article