Design and implementation of quality control plans that integrate moving average and internal quality control: incorporating the best of both worlds.
van Rossum, Huub H; van den Broek, Daan.
Afiliação
  • van Rossum HH; Department of Laboratory Medicine, The Netherlands Cancer Institute, Amsterdam, The Netherlands.
  • van den Broek D; Huvaros, Amsterdam, The Netherlands, Phone: +31-20-5122756, Fax: +31-20-5122799.
Clin Chem Lab Med ; 57(9): 1329-1338, 2019 08 27.
Article em En | MEDLINE | ID: mdl-30903753
Background New moving average quality control (MA QC) optimization methods have been developed and are available for laboratories. Having these methods will require a strategy to integrate MA QC and routine internal QC. Methods MA QC was considered only when the performance of the internal QC was limited. A flowchart was applied to determine, per test, whether MA QC should be considered. Next, MA QC was examined using the MA Generator (www.huvaros.com), and optimized MA QC procedures and corresponding MA validation charts were obtained. When a relevant systematic error was detectable within an average daily run, the MA QC was added to the QC plan. For further implementation of MA QC for continuous QC, MA QC management software was configured based on earlier proposed requirements. Also, protocols for the MA QC alarm work-up were designed to allow the detection of temporary assay failure based on previously described experiences. Results Based on the flowchart, 10 chemistry, two immunochemistry and six hematological tests were considered for MA QC. After obtaining optimal MA QC settings and the corresponding MA validation charts, the MA QC of albumin, bicarbonate, calcium, chloride, creatinine, glucose, magnesium, potassium, sodium, total protein, hematocrit, hemoglobin, MCH, MCHC, MCV and platelets were added to the QC plans. Conclusions The presented method allows the design and implementation of QC plans integrating MA QC for continuous QC when internal QC has limited performance.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Garantia da Qualidade dos Cuidados de Saúde / Gestão da Qualidade Total / Testes de Química Clínica Tipo de estudo: Sysrev_observational_studies Limite: Humans Idioma: En Ano de publicação: 2019 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Garantia da Qualidade dos Cuidados de Saúde / Gestão da Qualidade Total / Testes de Química Clínica Tipo de estudo: Sysrev_observational_studies Limite: Humans Idioma: En Ano de publicação: 2019 Tipo de documento: Article