Comparison between Roche Integra 400 plus and Abbott Architect ci8200 in Alanine aminotransferase assay.
Lamrabat, Somiya; Lyagoubi, Amina; Bouayadi, Ouardia; Elmalki, Jalila; Elkouche, Khadija; Bensalah, Mohammed; Choukri, Mohammed.
Afiliação
  • Lamrabat S; Central Laboratory of Mohammed VI University Hospital of Oujda City, Morocco.
  • Lyagoubi A; Faculty of Medicine and Pharmacy, First Mohammed University, Oujda City, Morocco.
  • Bouayadi O; Central Laboratory of Mohammed VI University Hospital of Oujda City, Morocco.
  • Elmalki J; Faculty of Medicine and Pharmacy, First Mohammed University, Oujda City, Morocco.
  • Elkouche K; Central Laboratory of Mohammed VI University Hospital of Oujda City, Morocco.
  • Bensalah M; Faculty of Medicine and Pharmacy, First Mohammed University, Oujda City, Morocco.
  • Choukri M; Central Laboratory of Mohammed VI University Hospital of Oujda City, Morocco.
Pract Lab Med ; 15: e00121, 2019 May.
Article em En | MEDLINE | ID: mdl-31011608
ABSTRACT

BACKGROUND:

The aim of this work is to present the results of a comparative study between the ALT assay on Integra 400 plus Roche Diagnostic versus Architect ci8200 of Abbott Diagnostic.

METHODS:

A total of 200 patients hospitalized in the various departments of the university hospital Mohammed VI of Oujda were prospectively tested on two systems Abbott Architect ci8200 and Roche Integra 400 plus. Both analyzers use the spectrophotometric technique by coupling the transamination reaction to the oxidation-reduction reaction at NAD. The agreement of the results between the different techniques was evaluated using the Bland-Altman difference diagram and the Passing-Bablok and Deming regression line.

RESULTS:

There was a high concordance between the two assays the equation of the Passing-Bablok line is YArchitect â€‹= â€‹-0,5625 â€‹+ â€‹0,9917 XIntegra with a correlation coefficient r2 = 0.999. The Bland-Altman diagram shows that the mean bias between the two methods is of the order of 1.1 IU/L and the difference between the ALT measurements by Architect and Integra is in the range of -1.4 to 3.6.

CONCLUSION:

Our study shows a high correlation of the ALT assay results between the architect ci8200 and Integra 400 plus.
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Ano de publicação: 2019 Tipo de documento: Article