Long-term results after standard endovascular aneurysm repair with the Endurant and Excluder stent grafts.
Oliveira-Pinto, José; Oliveira, Nelson F G; Bastos-Gonçalves, Frederico M; Hoeks, Sanne; Rijn, Marie Josee Van; Raa, Sander Ten; Mansilha, Armando; Verhagen, Hence J M.
Afiliação
  • Oliveira-Pinto J; Department of Vascular Surgery, Erasmus University Medical Centre, Rotterdam, The Netherlands; Department of Angiology and Vascular Surgery, Centro Hospitalar São João, Porto, Portugal; Department of Surgery and Physiology, Faculty of Medicine of Oporto, Porto, Portugal. Electronic address: oliveira
  • Oliveira NFG; Department of Vascular Surgery, Erasmus University Medical Centre, Rotterdam, The Netherlands; Department of Angiology and Vascular Surgery, Hospital do Divino Espírito Santo, Ponta Delgada, Azores, Portugal.
  • Bastos-Gonçalves FM; Department of Angiology and Vascular Surgery, Hospital de Santa Marta, Centro Hospitalar de Lisboa Central, Lisbon, Portugal; NOVA Medical School, Lisbon, Portugal.
  • Hoeks S; Department of Anesthesiology, Erasmus University Medical Center, Rotterdam, The Netherlands.
  • Rijn MJV; Department of Vascular Surgery, Erasmus University Medical Centre, Rotterdam, The Netherlands.
  • Raa ST; Department of Vascular Surgery, Erasmus University Medical Centre, Rotterdam, The Netherlands.
  • Mansilha A; Department of Angiology and Vascular Surgery, Centro Hospitalar São João, Porto, Portugal; Department of Surgery and Physiology, Faculty of Medicine of Oporto, Porto, Portugal.
  • Verhagen HJM; Department of Vascular Surgery, Erasmus University Medical Centre, Rotterdam, The Netherlands.
J Vasc Surg ; 71(1): 64-74, 2020 01.
Article em En | MEDLINE | ID: mdl-31147134
OBJECTIVE: Many endografts are currently available for standard endovascular repair of infrarenal abdominal aortic aneurysms. Comparison of long-term outcomes between devices might aid in this decision process, but comparative data are scarce. The purpose of this study was to report long-term clinical outcomes of two commercially available endoprosthesis, the Endurant (Medtronic Vascular, Inc, Minneapolis, Minn) and the Excluder (W. L. Gore & Associates, Flagstaff, Ariz) stent grafts. METHODS: Patients undergoing standard endovascular repair from July 2004 to December 2011 in a single institution with the Endurant or the Low-Porosity Excluder endografts were eligible. Only patients treated for intact degenerative abdominal infrarenal aneurysms were included. All measurements were performed on center-lumen line reconstructions obtained on dedicated software. The primary end point was primary clinical success, defined as clinical success without the need for an additional or secondary surgical or endovascular procedure. Neck-related events (a composite of type IA endoleak, neck-related secondary intervention, or migration of >5 mm), neck morphology changes, renal function, and overall survival were secondary end points. RESULTS: The study included 277 patients (156 Endurants; 121 Excluders). The median follow-up was 5.8 years (range, 0.1-12.4 years) and did not differ between groups (P = .18). Patients treated with the Endurant stent graft had wider (neck diameter of >28 mm, 27.3% vs 1.7% [P < .001]; neck diameter of 27 mm, [interquartile range (IQR), 24-29 mm] for Endurant and 24 mm [IQR, 22-25 mm] for Excluder; P < .001) and more angulated necks (ß-angle of >60°, 26.7% vs 12.5%; P = .004). Oversizing was greater in the Endurant group (16% [IQR, 12%-22%] vs 13% [IQR, 8%-17%], respectively; P < .001). Patients were treated outside device instructions for use regarding proximal neck: 16.7% in the Endurant and 17.3% in the Excluder group (P = .720). The 7-year primary clinical success was 54.7% for the Endurant and 58.1% for the Excluder groups (P = .53). Freedom from neck-related events at 7 years was 76.7% for the Endurant and 78.8% for Excluder group (P = .94). The Endurant stent graft (hazard ratio [HR], 2.7; 95% confidence interval [CI], 1.3-5.8; P = .009) was an independent predictor of significant renal function decline. Neck dilatation was greater in Endurant-implanted patients (13% [95% CI, 2%-22%] vs 4% [95% CI, 0%-10%]; P < .001). Overall survival at 7 years was 61.4% in the Endurant and 50.3% (n = 50; standard error, 0.047) in the Excluder group (P = .39). CONCLUSIONS: This study reveals that durable and sustainable results can be obtained with either of these late generation devices. This finding suggests that careful planning and a tailored device selection taking into account the patient's anatomy are more relevant determinants than the graft model itself to obtain clinical success. The Endurant endoprosthesis seems to be associated with a higher rate of neck dilatation and faster decrease in the estimated glomerular filtration rate, but further studies with longer follow-up are necessary to determine the clinical relevance of these findings.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Prótese Vascular / Stents / Aneurisma da Aorta Abdominal / Implante de Prótese Vascular / Procedimentos Endovasculares Tipo de estudo: Etiology_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Limite: Aged / Female / Humans / Male Idioma: En Ano de publicação: 2020 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Prótese Vascular / Stents / Aneurisma da Aorta Abdominal / Implante de Prótese Vascular / Procedimentos Endovasculares Tipo de estudo: Etiology_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Limite: Aged / Female / Humans / Male Idioma: En Ano de publicação: 2020 Tipo de documento: Article