Usefulness of a standard 12-lead electrocardiogram to predict the eligibility for a subcutaneous defibrillator.
Sakhi, Rafi; Theuns, Dominic A M J; Cosgun, Demet; Michels, Michelle; Schinkel, Arend F L; Kauling, R Martijn; Roos-Hesselink, Jolien W; Yap, Sing-Chien.
Afiliação
  • Sakhi R; Department of Cardiology, Erasmus MC, University Medical Center Rotterdam, Rotterdam, the Netherlands.
  • Theuns DAMJ; Department of Cardiology, Erasmus MC, University Medical Center Rotterdam, Rotterdam, the Netherlands.
  • Cosgun D; Department of Cardiology, Erasmus MC, University Medical Center Rotterdam, Rotterdam, the Netherlands.
  • Michels M; Department of Cardiology, Erasmus MC, University Medical Center Rotterdam, Rotterdam, the Netherlands.
  • Schinkel AFL; Department of Cardiology, Erasmus MC, University Medical Center Rotterdam, Rotterdam, the Netherlands.
  • Kauling RM; Department of Cardiology, Erasmus MC, University Medical Center Rotterdam, Rotterdam, the Netherlands.
  • Roos-Hesselink JW; Department of Cardiology, Erasmus MC, University Medical Center Rotterdam, Rotterdam, the Netherlands.
  • Yap SC; Department of Cardiology, Erasmus MC, University Medical Center Rotterdam, Rotterdam, the Netherlands. Electronic address: s.c.yap@erasmusmc.nl.
J Electrocardiol ; 55: 123-127, 2019.
Article em En | MEDLINE | ID: mdl-31170594
ABSTRACT

BACKGROUND:

Currently, the eligibility for a subcutaneous implantable defibrillator (S-ICD) system relies on a pre-implant vector screening based on the automated screening tool (AST). We investigated which 12-lead ECG characteristics are associated with eligibility for an S-ICD in a heterogeneous population at risk for sudden cardiac death (SCD). The goal is to determine patient eligibility for S-ICD using the standard 12-lead ECG, thereby avoiding additional AST screening.

METHODS:

We evaluated the eligibility for an S-ICD in 254 consecutive patients at risk for SCD. We identified 12-lead ECG parameters which were independently associated with AST passing (≥1 vector) using multivariable logistical regression analysis in our derivation cohort. The final model was tested in a separate validation cohort.

RESULTS:

The overall passing rate was 92% in our derivation cohort. Independent 12-lead ECG characteristics associated with AST passing were QRS ≤ 130 ms, absence of QRS/T discordance in lead II and R/T-ratio ≥3.5 in lead II. Eighty-three of 254 patients (33%) fulfilled these three criteria and had a passing rate of 100%. Of the validation cohort, 37 of 60 patients (62%) fulfilled all three criteria and also had a passing rate of 100%. The interobserver agreement for applying the ECG model was 90% (Cohen's Kappa = 0.80).

CONCLUSION:

Using the standard 12-lead ECG, we developed a simple screening model with a high specificity for S-ICD eligibility. Our results suggest that patients who fulfill the three ECG criteria do not need additional AST-screening.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Desfibriladores Implantáveis / Eletrocardiografia Tipo de estudo: Etiology_studies / Incidence_studies / Observational_studies / Prognostic_studies / Risk_factors_studies / Screening_studies Limite: Humans Idioma: En Ano de publicação: 2019 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Desfibriladores Implantáveis / Eletrocardiografia Tipo de estudo: Etiology_studies / Incidence_studies / Observational_studies / Prognostic_studies / Risk_factors_studies / Screening_studies Limite: Humans Idioma: En Ano de publicação: 2019 Tipo de documento: Article