Analytical evaluation and clinical application of insulin and C-peptide by a whole blood, lateral flow, point of care (POC) assay system.
Jekarl, Dong Wook; Choi, Hyunyu; Kim, Eun Sook; Lee, Seungok; Park, Hae-Il; Kim, Myungshin; Kim, Yonggoo.
Afiliação
  • Jekarl DW; a Department of Laboratory Medicine, The Catholic University of Korea, Incheon St. Mary's Hospital , Incheon , Korea.
  • Choi H; b Laboratory for Development and Evaluation Center, The Catholic University of Korea , Seoul , Korea.
  • Kim ES; a Department of Laboratory Medicine, The Catholic University of Korea, Incheon St. Mary's Hospital , Incheon , Korea.
  • Lee S; c Department of Internal Medicine, The Catholic University of Korea, Incheon St. Mary's Hospital , Incheon , Korea.
  • Park HI; a Department of Laboratory Medicine, The Catholic University of Korea, Incheon St. Mary's Hospital , Incheon , Korea.
  • Kim M; b Laboratory for Development and Evaluation Center, The Catholic University of Korea , Seoul , Korea.
  • Kim Y; b Laboratory for Development and Evaluation Center, The Catholic University of Korea , Seoul , Korea.
Scand J Clin Lab Invest ; 79(5): 347-353, 2019 Sep.
Article em En | MEDLINE | ID: mdl-31210062
ABSTRACT
The analytical performance and clinical application of measuring insulin and connecting peptide (C-peptide) by point of care (POC) assay were evaluated. A POC assay system (SelexOn, Osang Healthcare Inc., Anyang-si, Korea) was evaluated for precision, linearity, limit of blank (LOB), and limit of detection (LOD). Method comparison was performed with the Cobas Elecsys insulin and C-peptide assay (Roche Diagnostics GmbH, Mannheim, Germany) using 215 and 201 patient specimens for insulin and C-peptide, respectively. For clinical application, insulin resistance indices were studied. Homeostasis model assessment (HOMA) 1 and 2, Quantitative insulin sensitivity check index (QUIKI), fasting insulin resistance index (FIRI), and other indices were evaluated. The coefficient of variation (CV) of imprecision for low, medium, and high concentrations was 10.8%1, 15.99%, and 12.05%, respectively, for insulin and 9.21%, 13.51%, and 13.77%, respectively, for C-peptide. The linearity was validated to 839.78 pmol/L for insulin and to 17.30 nmol/L for C-peptide. LOB and LOD were 8.05 and 9.72 pmol/L for insulin and 0.05 and 0.08 nmol/L for C-peptide, respectively. For the method comparison, the regression equation was y = 1.259x - 8.818 (r = 0.957) for insulin and y = 1.163x - 0.088 (r = 0.985) for C-peptide. The ROC value and overall accuracy were as follows HOMA2 (C-peptide), 0.809, 79.7%; TyG, 0.788, 73.6%; CPR, 0.775, 74.8%; HOMA1, 0.725, 70.3%; QUIKI, 0.720, 70.3%; FIRI, 0.715, 70.1%; McAuley, 0.658, 65.1%; HOMA2 (Insulin), 0.645, 64.7%; Raynaud, 0.611, 61.4%, respectively. The POC assay system for insulin and C-peptide provided reliable results through a rapid and simple test that could be applied to clinical settings.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Bioensaio / Peptídeo C / Sistemas Automatizados de Assistência Junto ao Leito / Insulina Tipo de estudo: Evaluation_studies / Prognostic_studies Limite: Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2019 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Bioensaio / Peptídeo C / Sistemas Automatizados de Assistência Junto ao Leito / Insulina Tipo de estudo: Evaluation_studies / Prognostic_studies Limite: Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2019 Tipo de documento: Article