Improved HPLC-ECD analysis of mitomycin C, porfiromycin, VP 16-213 and VM 26 by implantation of software filters.
Ploegmakers, H H; Mertens, M J; van Oort, W J.
Afiliação
  • Ploegmakers HH; Department of Analytical Pharmacy, Faculty of Pharmacy, State University of Utrecht, The Netherlands.
Anticancer Res ; 7(6): 1315-9, 1987.
Article em En | MEDLINE | ID: mdl-3126702
ABSTRACT
In high performance liquid chromatography with electrochemical detection (HPLC/ECD) much attention has been paid to the performance of the applied detection system with respect to reliability and sensitivity. In general, insufficient attention has been paid to relatively easy to implant devices for improved collection and handling of detection signals. Four models of software filtering are studied and compared with hardware filtering. In the investigated chromatographic-electrochemical system, off-line parabolic filtering after on-line averaging of sixteen measurements proved to be the system of first choice, with respect to execution time, noise level, signal to noise ratio, peak height and resolution.