Agreement between supine and standing bioimpedance spectroscopy devices and dual-energy X-ray absorptiometry for body composition determination.
Esco, Michael R; Fedewa, Michael V; Freeborn, Todd J; Moon, Jordan R; Wingo, Jonathan E; Cicone, Zack; Holmes, Clifton J; Hornikel, Bjoern; Welborn, Bailey.
Afiliação
  • Esco MR; Department of Kinesiology, University of Alabama, Tuscaloosa, AL, USA.
  • Fedewa MV; Department of Kinesiology, University of Alabama, Tuscaloosa, AL, USA.
  • Freeborn TJ; Department of Electrical and Computer Engineering, University of Alabama, Tuscaloosa, AL, USA.
  • Moon JR; ImpediMed, Inc. Clinical Department, Carlsbad, CA, USA.
  • Wingo JE; Department of Kinesiology, University of Alabama, Tuscaloosa, AL, USA.
  • Cicone Z; Department of Kinesiology, University of Alabama, Tuscaloosa, AL, USA.
  • Holmes CJ; Department of Kinesiology, University of Alabama, Tuscaloosa, AL, USA.
  • Hornikel B; Department of Kinesiology, University of Alabama, Tuscaloosa, AL, USA.
  • Welborn B; Department of Kinesiology, University of Alabama, Tuscaloosa, AL, USA.
Clin Physiol Funct Imaging ; 39(5): 355-361, 2019 Sep.
Article em En | MEDLINE | ID: mdl-31287228
ABSTRACT

BACKGROUND:

Research comparing bioimpedance spectroscopy (BIS) to dual-energy X-ray absorptiometry (DXA) is limited, especially with newer BIS devices that take measures in a standing position instead of the traditional supine position.

PURPOSE:

The purpose of this study was to compare a standing BIS device (BISSTA ) and a supine BIS device (BISSUP ) to DXA for measuring body fat percentage (BF%), fat mass (FM) and fat-free mass (FFM) in a cohort of male and female subjects displaying a wide range of ages and BMI levels.

METHODS:

Ninety-five subjects (30 ± 15 years, 170 ± 8·0 cm, 72·6 ± 14·8 kg) participated in the study. Body composition measures were taken from BISSTA , BISSUP and DXA during a single visit to the laboratory following an 8- to 12-h fast in a euhydration state.

RESULTS:

Supine BIS device and BISSTA produced r-values >0·91 and low SEE values for all measurements compared to DXA. Effect sizes were 'trivial' for FFM comparing both BISSUP and BISSTA to DXA (<0·1) and 'small' for FM and BF% (<0·39). Compared to DXA, BISSTA resulted in lower total (TE) and constant errors/mean differences (CE) (TE < 3·6 kg, CE < -1·82 kg) versus BISSUP (TE < 4·35 kg, CE < -3·10 kg) for FFM.

CONCLUSION:

Fat-free mass values for BISSTA resulted in the most comparable measurements to DXA with no mean differences and the lowest total error and effect size. However, the findings indicated both BIS devices may be acceptable alternatives to DXA for BF%, FM and FFM in clinical practice.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Composição Corporal / Absorciometria de Fóton / Decúbito Dorsal / Posicionamento do Paciente / Posição Ortostática Tipo de estudo: Prognostic_studies Limite: Adolescent / Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2019 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Composição Corporal / Absorciometria de Fóton / Decúbito Dorsal / Posicionamento do Paciente / Posição Ortostática Tipo de estudo: Prognostic_studies Limite: Adolescent / Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2019 Tipo de documento: Article