Your browser doesn't support javascript.
loading
Retrospective assessment of a single fiducial marker tracking regimen with robotic stereotactic body radiation therapy for liver tumours.
Nakayama, Masao; Uehara, Kazuyuki; Nishimura, Hideki; Tamura, Shuhei; Munetomo, Yoshiki; Tsudou, Shinji; Mayahara, Hiroshi; Mukumoto, Naritoshi; Geso, Moshi; Sasaki, Ryohei.
Afiliação
  • Nakayama M; Division of Radiation Oncology, Kobe University Graduate School of Medicine, 7-5-2 Kusunokicho, Chuou-ku, Kobe City, Hyogo 650-0017, Japan.
  • Uehara K; Discipline of Medical Radiations, School of Biomedical & Health Sciences, RMIT University, Bundoora Campus, Victoria 3083, Australia.
  • Nishimura H; Division of Radiation Oncology, Kobe Minimally Invasive Cancer Center, 8-5-1 Minatojima-nakamachi, Chuou-ku, Kobe City, Hyogo 650-0046, Japan.
  • Tamura S; Division of Radiation Oncology, Kobe University Graduate School of Medicine, 7-5-2 Kusunokicho, Chuou-ku, Kobe City, Hyogo 650-0017, Japan.
  • Munetomo Y; Division of Radiation Oncology, Kobe Minimally Invasive Cancer Center, 8-5-1 Minatojima-nakamachi, Chuou-ku, Kobe City, Hyogo 650-0046, Japan.
  • Tsudou S; Division of Radiation Oncology, Kobe Minimally Invasive Cancer Center, 8-5-1 Minatojima-nakamachi, Chuou-ku, Kobe City, Hyogo 650-0046, Japan.
  • Mayahara H; Department of Radiation Oncology, Hyogo Cancer Center, 13-70 Kitaojicho, Akashi City, Hyogo 637-8558, Japan.
  • Mukumoto N; Division of Radiation Oncology, Kobe Minimally Invasive Cancer Center, 8-5-1 Minatojima-nakamachi, Chuou-ku, Kobe City, Hyogo 650-0046, Japan.
  • Geso M; Division of Radiation Oncology, Kobe University Graduate School of Medicine, 7-5-2 Kusunokicho, Chuou-ku, Kobe City, Hyogo 650-0017, Japan.
  • Sasaki R; Discipline of Medical Radiations, School of Biomedical & Health Sciences, RMIT University, Bundoora Campus, Victoria 3083, Australia.
Rep Pract Oncol Radiother ; 24(4): 383-391, 2019.
Article em En | MEDLINE | ID: mdl-31297039
ABSTRACT

AIM:

To investigate tumour motion tracking uncertainties in the CyberKnife Synchrony system with single fiducial marker in liver tumours.

BACKGROUND:

In the fiducial-based CyberKnife real-time tumour motion tracking system, multiple fiducial markers are generally used to enable translation and rotation corrections during tracking. However, sometimes a single fiducial marker is employed when rotation corrections are not estimated during treatment. MATERIALS AND

METHODS:

Data were analysed for 32 patients with liver tumours where one fiducial marker was implanted. Four-dimensional computed tomography (CT) scans were performed to determine the internal target volume (ITV). Before the first treatment fraction, the CT scans were repeated and the marker migration was determined. Log files generated by the Synchrony system were obtained after each treatment and the correlation model errors were calculated. Intra-fractional spine rotations were examined on the spine alignment images before and after each treatment.

RESULTS:

The mean (standard deviation) ITV margin was 4.1 (2.3) mm, which correlated weakly with the distance between the fiducial marker and the tumour. The mean migration distance of the marker was 1.5 (0.7) mm. The overall mean correlation model error was 1.03 (0.37) mm in the radial direction. The overall mean spine rotations were 0.27° (0.31), 0.25° (0.22), and 0.23° (0.26) for roll, pitch, and yaw, respectively. The treatment time was moderately associated with the correlation model errors and weakly related to spine rotation in the roll and yaw planes.

CONCLUSIONS:

More caution and an additional safety margins are required when tracking a single fiducial marker.
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Tipo de estudo: Risk_factors_studies Idioma: En Ano de publicação: 2019 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Tipo de estudo: Risk_factors_studies Idioma: En Ano de publicação: 2019 Tipo de documento: Article