Your browser doesn't support javascript.
loading
An inappropriate shock after reprogramming tachycardia zones.
Marinheiro, Rita; Parreira, Leonor; Amador, Pedro; Mesquita, Dinis; Almeida, Lurdes; Lousinha, Ana; Farinha, José; Fonseca, Marta; Esteves, Ana Fátima; Caria, Rui.
Afiliação
  • Marinheiro R; Centro Hospitalar de Setubal EPE Setubal Portugal.
  • Parreira L; Centro Hospitalar de Setubal EPE Setubal Portugal.
  • Amador P; Centro Hospitalar de Setubal EPE Setubal Portugal.
  • Mesquita D; Centro Hospitalar de Setubal EPE Setubal Portugal.
  • Almeida L; Centro Hospitalar Barreiro-Montijo EPE Barreiro Portugal.
  • Lousinha A; Centro Hospitalar Lisboa Central EPE Lisboa Portugal.
  • Farinha J; Centro Hospitalar de Setubal EPE Setubal Portugal.
  • Fonseca M; Centro Hospitalar de Setubal EPE Setubal Portugal.
  • Esteves AF; Centro Hospitalar de Setubal EPE Setubal Portugal.
  • Caria R; Centro Hospitalar de Setubal EPE Setubal Portugal.
J Arrhythm ; 35(4): 679-681, 2019 Aug.
Article em En | MEDLINE | ID: mdl-31410241
ABSTRACT
A patient with an implantable cardioverter-defibrillator (ICD) (Abbott®) had episodes of slow monomorphic ventricular tachycardia (VT) and his ICD was programmed with three tachycardia zones. During the follow-up, he received an inappropriate shock. Upon interrogation (of the device), trigeminal pattern binned as ventricular sensing (VS)-VS-ventricular fibrillation (VF) was detected. VF was assumed according to binning system. When VF is present, discrimination algorithms are not available and five consecutive sinus beats are necessary to reset binning system. Catheter ablation was performed to treat VT in order to reprogram tachycardia zones.
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Ano de publicação: 2019 Tipo de documento: Article