Your browser doesn't support javascript.
loading
Linezolid Resistance in Methicillin-Resistant Staphylococcus aureus in Korea: High Rate of False Resistance to Linezolid by the VITEK 2 System.
Yoo, In Young; Kang, On Kyun; Shim, Hyang Jin; Huh, Hee Jae; Lee, Nam Yong.
Afiliação
  • Yoo IY; Department of Laboratory Medicine and Genetics, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Korea.
  • Kang OK; Department of Laboratory Medicine and Genetics, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Korea.
  • Shim HJ; Center for Clinical Medicine, Samsung Biomedical Research Institute, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Korea.
  • Huh HJ; Department of Laboratory Medicine and Genetics, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Korea. pmhhj77@gmail.com.
  • Lee NY; Department of Laboratory Medicine and Genetics, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Korea. micro.lee@samsung.com.
Ann Lab Med ; 40(1): 57-62, 2020 Jan.
Article em En | MEDLINE | ID: mdl-31432640
ABSTRACT
As various linezolid resistance mechanisms have been identified in methicillin-resistant Staphylococcus aureus (MRSA), we investigated the molecular characteristics of MRSA with elevated linezolid minimum inhibitory concentrations (MICs), using the VITEK 2 system (bioMérieux, Marcy-l'Étoile, France). Twenty-seven MRSA isolates from 14 patients exhibiting linezolid MICs ≥8 µg/mL were examined by broth microdilution (BMD) test as well as by sequencing for mutations in the 23S rRNA gene or ribosomal proteins (L3, L4, and L22) and the presence of the optrA, cfr, and cfr(B) genes. Of the 27 isolates, four (14.8%) from one patient were confirmed as linezolid resistant by BMD and harbored a 23S rRNA T2500A mutation. The remaining 23 were confirmed as linezolid susceptible, indicating that the linezolid-resistant results were major errors generated by VITEK 2. The most commonly detected mutation (19/27, 70.4%), L3 Gly152Asp, was detected in only linezolid-susceptible isolates. No isolates contained optrA, cfr, or cfr(B) or any L4 or L22 protein alterations. Our results show that the 23S rRNA T2500A mutation was mainly associated with linezolid resistance, while the L3 Gly152Asp mutation was not related to linezolid resistance. A confirmatory test is recommended for VITEK 2 linezolid-resistant results owing to the high probability of false resistant results.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Testes de Sensibilidade Microbiana / Farmacorresistência Bacteriana / Staphylococcus aureus Resistente à Meticilina / Linezolida Tipo de estudo: Prognostic_studies Limite: Humans País/Região como assunto: Asia Idioma: En Ano de publicação: 2020 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Testes de Sensibilidade Microbiana / Farmacorresistência Bacteriana / Staphylococcus aureus Resistente à Meticilina / Linezolida Tipo de estudo: Prognostic_studies Limite: Humans País/Região como assunto: Asia Idioma: En Ano de publicação: 2020 Tipo de documento: Article