Your browser doesn't support javascript.
loading
Dimensions of artefacts caused by cochlear and auditory brainstem implants in magnetic resonance imaging.
Majdani, Elham; Majdani, Omid; Steffens, Melanie; Warnecke, Athanasia; Lesinski-Schiedat, Anke; Lenarz, Thomas; Götz, Friedrich.
Afiliação
  • Majdani E; Department of Otolaryngology, Hannover Medical School, Hannover, Germany.
  • Majdani O; Department of Otolaryngology, Hannover Medical School, Hannover, Germany.
  • Steffens M; Cluster of Excellence of the German Research Foundation (DFG; 'Deutsche Forschungsgemeinschaft') 'Hearing4all', Hannover, Germany.
  • Warnecke A; Department of Otolaryngology, Hannover Medical School, Hannover, Germany.
  • Lesinski-Schiedat A; Department of Otolaryngology, Hannover Medical School, Hannover, Germany.
  • Lenarz T; Cluster of Excellence of the German Research Foundation (DFG; 'Deutsche Forschungsgemeinschaft') 'Hearing4all', Hannover, Germany.
  • Götz F; Department of Otolaryngology, Hannover Medical School, Hannover, Germany.
Cochlear Implants Int ; 21(2): 67-74, 2020 03.
Article em En | MEDLINE | ID: mdl-31553273
ABSTRACT

Background:

The aim of the study was to investigate the extent of MRI artefacts due to the magnet of selected auditory implants.Study

design:

Artefacts of the Synchrony cochlear implant at 1.5 T as well as at 3 T MRI devices were examined in cadavers and compared to the artefacts in MRI scans at 1.5 T of 17 patients implanted with CI (n = 12) and auditory brainstem implants (ABI) (n = 5).

Results:

None of the scanned implants showed any failure after MRI. After removal of the magnet, only a portion of the images in the direct neighbourhood of the implant, especially in the temporal and parietal lobe, contained artefacts. More anatomical substructures were visible without artefacts using the MedEl Synchrony device.

Conclusion:

Artefacts can be markedly reduced by rotating, self-aligning magnet. Removal of the magnet also results in reduction of artefacts.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Imageamento por Ressonância Magnética / Implantes Cocleares / Artefatos / Implantes Auditivos de Tronco Encefálico / Perda Auditiva Neurossensorial Tipo de estudo: Evaluation_studies / Observational_studies Limite: Humans Idioma: En Ano de publicação: 2020 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Imageamento por Ressonância Magnética / Implantes Cocleares / Artefatos / Implantes Auditivos de Tronco Encefálico / Perda Auditiva Neurossensorial Tipo de estudo: Evaluation_studies / Observational_studies Limite: Humans Idioma: En Ano de publicação: 2020 Tipo de documento: Article