Your browser doesn't support javascript.
loading
Characterization of Corneal Biomechanical Properties and Determination of Natural Intraocular Pressure Using CID-GAT.
Lu, Shu-Hao; Chong, I T; Leung, Stanley Y Y; Lam, David C C.
Afiliação
  • Lu SH; 1Department of Mechanical and Aerospace Engineering, The Hong Kong University of Science and Technology, Hong Kong, China.
  • Chong IT; 1Department of Mechanical and Aerospace Engineering, The Hong Kong University of Science and Technology, Hong Kong, China.
  • Leung SYY; 1Department of Mechanical and Aerospace Engineering, The Hong Kong University of Science and Technology, Hong Kong, China.
  • Lam DCC; 1Department of Mechanical and Aerospace Engineering, The Hong Kong University of Science and Technology, Hong Kong, China.
Transl Vis Sci Technol ; 8(5): 10, 2019 Sep.
Article em En | MEDLINE | ID: mdl-31579556
ABSTRACT

PURPOSE:

The intraocular pressure (IOP) measured using Goldmann Applanation Tonometry (GAT) is confounded by individual corneal properties. We investigated a modified method that removes the confoundment by incorporating corneal properties into the Imbert-Fick's law is investigated.

METHOD:

Porcine eyes were pressurized between 10 and 40 mm Hg using a manometer. The eyes were indented using a flat cylindrical indenter. A modified corneal indentation device (CID) procedure was used to obtain the corneal moduli Eqs . The calculated IOPNC from the Imbert-Fick's Law using the corneal moduli Eqs was compared to the natural IOPN, measured using pressure sensor inserted into the eye.

RESULTS:

Test results showed that IOP-dependent corneal modulus Eqs is a primary confounding factor in IOP calculation. The average elastic modulus Eqs is 0.173 ± 0.018 MPa at 20 mm Hg, and increases with IOP at a linear rate of 0.0066 MPa per mm Hg (r = 0.997, P < 0.001). Incorporation of individual Eqs into IOPNC calculation showed that IOPNC are in good agreement with reference IOPN (slope = 0.999, r = 0.939, P < 0.001).

CONCLUSIONS:

The IOP-dependent corneal modulus Eqs is a primary confounding factor in IOP calculation. A modified CID-GAT procedure to obtain natural cornea-independent IOPNC is developed and verified in this study. The CID-GAT IOP modification may be used in place of conventional GAT when the confounding effects in eyes with atypical cornea (e.g., laser-assisted in situ keratomileusis [LASIK] thinned) are significant. TRANSLATIONAL RELEVANCE Confoundment from corneal properties results in IOP measurement errors. The study showed that the CID-GAT method can significantly reduce the confounding corneal errors.