Assessing the limit of detection of Fourier-transform infrared spectroscopy and immunoassay strips for fentanyl in a real-world setting.
McCrae, Karen; Tobias, Samuel; Grant, Cameron; Lysyshyn, Mark; Laing, Richard; Wood, Evan; Ti, Lianping.
Afiliação
  • McCrae K; British Columbia Centre on Substance Use, Vancouver, Canada.
  • Tobias S; British Columbia Centre on Substance Use, Vancouver, Canada.
  • Grant C; British Columbia Centre on Substance Use, Vancouver, Canada.
  • Lysyshyn M; Vancouver Coastal Health, Vancouver, Canada.
  • Laing R; School of Population and Public Health, University of British Columbia, Vancouver, Canada.
  • Wood E; Health Canada, Drug Analysis Service, Burnaby, Canada.
  • Ti L; British Columbia Centre on Substance Use, Vancouver, Canada.
Drug Alcohol Rev ; 39(1): 98-102, 2020 01.
Article em En | MEDLINE | ID: mdl-31746056
ABSTRACT
INTRODUCTION AND

AIMS:

Drug checking is a harm reduction intervention increasingly used in the context of the opioid overdose epidemic. The aim of the study was to determine the limit of detection for fentanyl of two point-of-care drug checking technologies. DESIGN AND

METHODS:

Samples tested at point-of-care using Bruker Fourier transform infrared (FTIR) spectroscopy and BTNX fentanyl immunoassay strips were sent for confirmatory laboratory analysis using quantitative nuclear magnetic resonance (qNMR) spectroscopy. Concentrations by weight were determined and compared to results obtained with point-of-care methods.

RESULTS:

In total, 283 samples were sent for qNMR analysis; among these, 173 (61.1%) tested positive for fentanyl. As determined by qNMR, fentanyl concentration by weight ranged from 1% to 91%. Among these 173 samples, fentanyl was not detected in 30 (17.3%) samples by FTIR and in 4 (2.3%) samples by test strip. Samples containing fentanyl that went undetected by FTIR had concentrations ≤10%. The four samples containing fentanyl that went undetected by test strip had concentrations ≤5% (i.e. 1%, 3%, 4%, 5%). DISCUSSION AND

CONCLUSIONS:

Fentanyl immunoassay strips were able to consistently detect the presence of fentanyl in samples at lower concentrations than FTIR spectroscopy. Given that FTIR spectroscopy is able to quantify content, mixture and concentrations on an array of compounds beyond just fentanyl but requires concentrations generally greater than 10%, these findings provide evidence for use of FTIR spectroscopy and immunoassay strips in combination to compensate for the limitations of each technology alone.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Imunoensaio / Fentanila / Espectroscopia de Infravermelho com Transformada de Fourier / Limite de Detecção / Analgésicos Opioides Tipo de estudo: Diagnostic_studies Limite: Humans Idioma: En Ano de publicação: 2020 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Imunoensaio / Fentanila / Espectroscopia de Infravermelho com Transformada de Fourier / Limite de Detecção / Analgésicos Opioides Tipo de estudo: Diagnostic_studies Limite: Humans Idioma: En Ano de publicação: 2020 Tipo de documento: Article