Development of a dynamic blood vessel phantom for evaluation of moving images.
Osawa, Miwa; Kondo, Keisuke; Hisaya, Sato; Kato, Kyoichi.
Afiliação
  • Osawa M; Graduate School of Medical Health Sciences, Komazawa University, Setagaya-ku, Tokyo, Japan.
  • Kondo K; Radiation Technology Department, Showa University Fujigaoka Hospital, Aoba-ku, Yokohama city, Kanagawa, Japan.
  • Hisaya S; Graduate School of Medical Health Sciences, Komazawa University, Setagaya-ku, Tokyo, Japan.
  • Kato K; Graduate School of Health Sciences, Showa University, Shinagawa-ku, Tokyo, Japan.
J Appl Clin Med Phys ; 20(12): 193-203, 2019 Dec.
Article em En | MEDLINE | ID: mdl-31778015
ABSTRACT
In coronary angiography (CAG) and percutaneous coronary intervention (PCI), it is important for radiological technologists to optimize the balance between radiation dose and image quality for physicians to be able to perform CAG and PCI most effectively. Evaluation of image processing is necessary to ensure that technologists can optimally adjust image quality for clinical use to the extent that physicians require. However, few phantoms are available for evaluating fluoroscopic image processing, and this makes it necessary to adjust image quality in clinical settings while utilizing the image processing according to the manufacturer's recommendations. In this study, we developed a dynamic phantom that mimics a pulsating coronary artery for use in image quality analysis of moving images. We also examined whether processed images (image lag) can be physically analyzed. Two issues require special attention in creating a new phantom establishing the exact position of the simulated blood vessel in the phantom, and providing good reproducibility. The study used the rotational motion of a disk to generate images, with a circular simulated blood vessel on the rotating acrylic disk, with the center of the simulated blood vessel shifted from the center of the acrylic disk. This enabled the reproduction of translational motion of the simulated blood vessel. As a result, because this phantom has signals and afterimages at the same position (of a simulated blood vessel), a quantitative evaluation of the afterimages became possible. In the evaluation of the image processing with the angiograph, it was shown that evaluations of image lag which are typically performed in clinical settings can be performed with the dynamic blood vessel phantom.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Vasos Sanguíneos / Angiografia Coronária / Imagens de Fantasmas / Vasos Coronários Tipo de estudo: Guideline Limite: Humans Idioma: En Ano de publicação: 2019 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Vasos Sanguíneos / Angiografia Coronária / Imagens de Fantasmas / Vasos Coronários Tipo de estudo: Guideline Limite: Humans Idioma: En Ano de publicação: 2019 Tipo de documento: Article