Predictors of successful ultrasound-guided lead implantation.
Ahmed, Asim S; Gilge, Jasen L; Clark, Brad A; Shah, Ankur; Bagga, Shiv; Padanilam, Mathew S; Prystowsky, Eric N; Joshi, Sandeep A; Nair, Girish V; Patel, Parin J.
Afiliação
  • Ahmed AS; St. Vincent Medical Group, St. Vincent Hospital, Indianapolis, Indiana.
  • Gilge JL; St. Vincent Medical Group, St. Vincent Hospital, Indianapolis, Indiana.
  • Clark BA; St. Vincent Medical Group, St. Vincent Hospital, Indianapolis, Indiana.
  • Shah A; Doctors Hospital-OhioHealth, Columbus, Ohio.
  • Bagga S; St. Vincent Medical Group, St. Vincent Hospital, Indianapolis, Indiana.
  • Padanilam MS; St. Vincent Medical Group, St. Vincent Hospital, Indianapolis, Indiana.
  • Prystowsky EN; St. Vincent Medical Group, St. Vincent Hospital, Indianapolis, Indiana.
  • Joshi SA; St. Vincent Medical Group, St. Vincent Hospital, Indianapolis, Indiana.
  • Nair GV; St. Vincent Medical Group, St. Vincent Hospital, Indianapolis, Indiana.
  • Patel PJ; St. Vincent Medical Group, St. Vincent Hospital, Indianapolis, Indiana.
Pacing Clin Electrophysiol ; 43(2): 217-222, 2020 02.
Article em En | MEDLINE | ID: mdl-31808167
ABSTRACT

BACKGROUND:

Technical advances have improved the safety of cardiac implantable electronic device (CIED) insertion, but periprocedural complications persist. Despite ultrasound (US) guidance for vascular access being feasible and exhibiting shorter fluoroscopy times, it is not widely adopted for insertion of CIEDs. Thus, we studied the use of US for CIED insertion to (1) quantify the success rate of venous cannulation, (2) identify predictors of failed cannulation, and (3) quantify the rate of complications using US guidance.

METHODS:

We studied 166 consecutive patients who underwent US-guided CIED implantation. Anatomic parameters of the axillary vein were measured. The primary outcome was success (group 1) or failure (group 2) to obtain vascular access utilizing US guidance. Secondary outcomes included pneumothorax and hematoma.

RESULTS:

Successful US-guided cannulation occurred in 154 of 166 patients (93%). No patient had a pneumothorax. Hematoma occurred in 1 of 166 patients (0.01%). Group 2 exhibited higher male proportion at 11 of 12 (92%) compared with 94 of 154 (61%) in group 1 (P = .03), increased vein depth at 3.84 versus 2.85 cm (P = .003), more right-sided implants (P = .03), higher weight at 104.6 versus 85.3 kg (P = .017), higher body mass index at 35.6 versus 29.2 kg/m2 (P = .049), and higher body surface area at 2.24 versus 1.99 m2 (P = .013). Other parameters were statistically nonsignificant. In multivariate analysis, vein depth remained significantly associated with failure.

CONCLUSION:

Using US guidance for CIED implantation is successful in the vast majority (93%) of patients. Rare cases of unsuccessful cannulation were associated with right-sided implants and increased venous depth.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Desfibriladores Implantáveis / Ultrassonografia de Intervenção / Implantação de Prótese Tipo de estudo: Observational_studies / Prognostic_studies / Risk_factors_studies Limite: Aged / Female / Humans / Male Idioma: En Ano de publicação: 2020 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Desfibriladores Implantáveis / Ultrassonografia de Intervenção / Implantação de Prótese Tipo de estudo: Observational_studies / Prognostic_studies / Risk_factors_studies Limite: Aged / Female / Humans / Male Idioma: En Ano de publicação: 2020 Tipo de documento: Article