Validation of the Spacelabs 90227 OnTrak device according to the European and British Hypertension Societies as well as the American protocols.
de Greeff, Annemarie; Shennan, Andrew H.
Afiliação
  • de Greeff A; Accuracy Assessed Medical Devices CC, Kimberley, South Africa.
  • Shennan AH; Department of Women and Children's Health, School of Health Course Sciences, King's College London, London, UK.
Blood Press Monit ; 25(2): 110-114, 2020 Apr.
Article em En | MEDLINE | ID: mdl-31833949
ABSTRACT

OBJECTIVE:

To assess the static device accuracy of the Spacelabs 90227 OnTrak according to the three most recognized validation protocols, in both adults and children.

METHODS:

Subjects were recruited and data gathered and analysed according to the European Society of Hypertension (ESH-IP2), the British Hypertension Society (BHS) and the American National Standards Institute/Association for the Advancement of Medical Instrumentation/International Standards Organization (ANSI/AAMI/ISO) protocols.

RESULTS:

The device achieved a mean difference ± SD between observers and device for SBP/DBP was 0.9 ± 7/-1.4 ± 6 mmHg (ESH-IP2) and 1.7 ± 8/-1.5/6 mmHg (BHS) Grade A/A. For the ANSI/AAMI/ISO protocol, the average device-observer difference for SBP/DBP was 0.4 ± 7/-1.5 ± 7 mmHg fulfilling the two criteria of the protocol.

CONCLUSION:

The Spacelabs 90227 OnTrak device achieved the criteria of all three recognised validation protocols and can therefore be recommended for clinical use as a static device.