Ultra-focal Magnetic Stimulation Using a µTMS coil: a Computational Study.
Annu Int Conf IEEE Eng Med Biol Soc ; 2019: 3987-3990, 2019 Jul.
Article em En | MEDLINE | ID: mdl-31946745
ABSTRACT
A new miniaturized figure-of-eight coil (µCoil) for TMS applications has been developed taking advantage of the Flex circuit technology. First experiments on volunteers demonstrated the ability of the µCoil to elicit sensorial action potentials of the peripheral nervous system. The necessity of reducing the size of standard TMS stimulator arises from the poor spatial resolution of the latter. This study aims to model the µCoil and study the electromagnetic fields induced inside the arm during peripheral nerve stimulation. Results confirmed that the µCoil is capable of inducing a focalized electric field inside the tissues with amplitudes up to 70V/m consistent with the observed peripheral nerve stimulation in healthy volunteers.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Estimulação Magnética Transcraniana / Magnetismo Limite: Humans Idioma: En Ano de publicação: 2019 Tipo de documento: Article