Wide-field time-gated SPAD imager for phasor-based FLIM applications.
Methods Appl Fluoresc
; 8(2): 024002, 2020 Feb 05.
Article
em En
| MEDLINE
| ID: mdl-31968310
ABSTRACT
We describe the performance of a new wide area time-gated single-photon avalanche diode (SPAD) array for phasor-FLIM, exploring the effect of gate length, gate number and signal intensity on the measured lifetime accuracy and precision. We conclude that the detector functions essentially as an ideal shot noise limited sensor and is capable of video rate FLIM measurement. The phasor approach used in this work appears ideally suited to handle the large amount of data generated by this type of very large sensor (512 × 512 pixels), even in the case of small number of gates and limited photon budget.
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Idioma:
En
Ano de publicação:
2020
Tipo de documento:
Article