Influence of tumor shape and location in eye plaque brachytherapy dosimetry.
Studenski, Matthew T; Patel, Nirav V; Markoe, Arnold; Harbour, J William; Samuels, Stuart E.
Afiliação
  • Studenski MT; Department of Radiation Oncology, University of Miami Sylvester Comprehensive Cancer Center, Miami, FL. Electronic address: m.studenski@med.miami.edu.
  • Patel NV; Department of Radiation Oncology, University of Miami Sylvester Comprehensive Cancer Center, Miami, FL.
  • Markoe A; Department of Radiation Oncology, University of Miami Sylvester Comprehensive Cancer Center, Miami, FL.
  • Harbour JW; Department of Ophthalmology, University of Miami Bascom Palmer Eye Institute, Miami, FL.
  • Samuels SE; Department of Radiation Oncology, University of Miami Sylvester Comprehensive Cancer Center, Miami, FL.
Brachytherapy ; 19(2): 249-254, 2020.
Article em En | MEDLINE | ID: mdl-32061535
ABSTRACT

PURPOSE:

A common treatment planning technique for eye plaque brachytherapy is to model the tumor as an ellipse. For posterior tumors near the optic disc and fovea, this approach may lead to overlap between tumor and the organ at risk (OAR). We hypothesized that a superior plan can be generated by modeling the actual tumor shape. MATERIALS AND

METHODS:

Forty eye plaque patients with tumors <1 cm from the optic disc and fovea were selected. Two treatment plans were generated for each patient an elliptical tumor model plan and a true tumor model plan. Dosimetric data were collected for each plan, and Wilcoxon signed-rank tests were used to asses any statistically significant differences.

RESULTS:

Equivalent tumor coverage was confirmed between the elliptical and true tumor plans for all patients. Qualitative analysis showed greater dosimetric differences between plans as the distance from the OARs increased from 0 to 2 mm but the largest differences were observed between 2 and 4 mm. Minimal differences between models were seen beyond 4 mm. Statistically significant dosimetric improvements were found for tumors <4 mm from the fovea and <2 mm from the optic disc.

CONCLUSIONS:

Intuitively, accurate modeling of the tumor accounting for irregularities in the shape should result in a more conformal plan and an overall reduction in OAR dose. However, this technique is only beneficial for tumors that are within 4 mm of the fovea or optic disc. An elliptical tumor model allows for an acceptable plan unless the tumor is located posteriorly and has an irregular shape.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Planejamento da Radioterapia Assistida por Computador / Braquiterapia / Neoplasias Oculares / Melanoma Tipo de estudo: Etiology_studies / Qualitative_research Limite: Humans Idioma: En Ano de publicação: 2020 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Planejamento da Radioterapia Assistida por Computador / Braquiterapia / Neoplasias Oculares / Melanoma Tipo de estudo: Etiology_studies / Qualitative_research Limite: Humans Idioma: En Ano de publicação: 2020 Tipo de documento: Article