Design and clinical validation of a software program for automated measurement of mammographic breast density.
Araújo, Adriano L C; Soares, Heliana B; Carvalho, Daniel F; Mendonça, Roberto M; Oliveira, Antonio G.
Afiliação
  • Araújo ALC; Department of Radiology, Hospital Universitário Onofre Lopes, Universidade Federal do Rio Grande do Norte, Av. Nilo Peçanha 620, Petrópolis, Natal, RN, 59012-300, Brazil. adrianorad@gmail.com.
  • Soares HB; Instituto de Radiologia de Natal, Av. Afonso Pena 744 - Tirol, Natal, RN, 59020-100, Brazil. adrianorad@gmail.com.
  • Carvalho DF; Department of Biomedical Engineering, Centro de Tecnologia, Universidade Federal do Rio Grande do Norte, Campus Universitário, Av. Senador Salgado Filho 300, Lagoa Nova, Natal, RN, 59078-970, Brazil.
  • Mendonça RM; Department of Biomedical Engineering, Centro de Tecnologia, Universidade Federal do Rio Grande do Norte, Campus Universitário, Av. Senador Salgado Filho 300, Lagoa Nova, Natal, RN, 59078-970, Brazil.
  • Oliveira AG; Department of Radiology, Hospital Universitário Onofre Lopes, Universidade Federal do Rio Grande do Norte, Av. Nilo Peçanha 620, Petrópolis, Natal, RN, 59012-300, Brazil.
BMC Med Inform Decis Mak ; 20(1): 45, 2020 03 02.
Article em En | MEDLINE | ID: mdl-32122371
BACKGROUND: Mammographic breast density is an important predictor of breast cancer, but its measurement has limitations related to subjectivity of visual evaluation or to difficult access for automatic volumetric measurement methods. Herein, we describe the design and clinical validation of Aguida, a software program for automated quantification of breast density from flat mammography images. MATERIALS AND METHODS: The software program was developed in MatLab. After image segmentation separating the background from the breast image, the operator positions a cursor defining a region of interest on the pectoralis major muscle from the mediolateral oblique view. Then, in the craniocaudal view, the threshold for separation of the dense tissue is based on the optical density of the pectoral muscle, and the proportion of dense tissue is calculated by the program. Mammograms obtained from 2 different occasions in 291 women were used for clinical evaluation. RESULTS: The intraclass correlation coefficient (ICC) between breast density measurements by the software and by a radiologist was 0.96, with a bias of only 0.67 percentage points and a 95% limit of agreement of 13.5 percentage points; the ICC was 0.94 in the interobserver reliability assessment by two radiologists with different experience; and the ICC was 0.98 in the intraobserver reliability assessment. The distribution among the density classes was close to the values obtained with the volumetric software. CONCLUSIONS: Measurement of breast density with the Aguida program from flat mammography images showed high agreement with the visual determination by radiologists, and high inter- and intra-observer reliability.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Validação de Programas de Computador / Neoplasias da Mama / Mamografia / Interpretação de Imagem Radiográfica Assistida por Computador / Densidade da Mama Tipo de estudo: Prognostic_studies Limite: Adult / Aged / Aged80 / Female / Humans / Middle aged Idioma: En Ano de publicação: 2020 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Validação de Programas de Computador / Neoplasias da Mama / Mamografia / Interpretação de Imagem Radiográfica Assistida por Computador / Densidade da Mama Tipo de estudo: Prognostic_studies Limite: Adult / Aged / Aged80 / Female / Humans / Middle aged Idioma: En Ano de publicação: 2020 Tipo de documento: Article