Failure at the femoral stem extension-condylar interface in a rotating hinge knee without radiographic evidence of loosening.
Arthroplast Today
; 6(1): 36-40, 2020 Mar.
Article
em En
| MEDLINE
| ID: mdl-32211472
Modular stem extensions have become ubiquitous in revision total knee arthroplasty systems. Although stem extensions are valuable in addressing bone deficiencies and improving implant fixation, the stem extension-condylar interface may be a point of implant failure. We report a case of failure at the femoral stem extension-condylar interface in a Zimmer NexGen Rotating Hinge Knee (Zimmer, Warsaw, IN). Currently, several published case reports describe failure at the femoral stem extension-condylar interface but only 1 case describes loosening at a taper junction without evidence of set screw failure or taper fracture. Furthermore, no published cases describe this type of failure in the Zimmer NexGen Rotating Hinge Knee (Zimmer).