Location of Reentry Tears Affects False Lumen Thrombosis in Aortic Dissection Following TEVAR.
Armour, Chlöe Harriet; Menichini, Claudia; Milinis, Kristijonas; Gibbs, Richard G J; Xu, Xiao Yun.
Afiliação
  • Armour CH; Department of Chemical Engineering, Imperial College London, UK.
  • Menichini C; Department of Chemical Engineering, Imperial College London, UK.
  • Milinis K; Regional Vascular Unit, St Mary's Hospital, Imperial College Healthcare NHS Trust, London, UK.
  • Gibbs RGJ; Regional Vascular Unit, St Mary's Hospital, Imperial College Healthcare NHS Trust, London, UK.
  • Xu XY; Department of Chemical Engineering, Imperial College London, UK.
J Endovasc Ther ; 27(3): 396-404, 2020 Jun.
Article em En | MEDLINE | ID: mdl-32364001
ABSTRACT

Purpose:

To report a study that assesses the influence of the distance between the distal end of a thoracic stent-graft and the first reentry tear (SG-FRT) on the progression of false lumen (FL) thrombosis in patients who underwent thoracic endovascular aortic repair (TEVAR). Materials and

Methods:

Three patient-specific geometrical models were reconstructed from postoperative computed tomography scans. Two additional models were created by artificially changing the SG-FRT distance in patients 1 and 2. In all 5 models, computational fluid dynamics simulations coupled with thrombus formation modeling were performed at physiological flow conditions. Predicted FL thrombosis was compared to follow-up scans.

Results:

There was reduced false lumen flow and low time-averaged wall shear stress (TAWSS) in patients with large SG-FRT distances. Predicted thrombus formation and growth were consistent with follow-up scans for all patients. Reducing the SG-FRT distance by 30 mm in patient 1 increased the flow and time-averaged wall shear stress in the upper abdominal FL, reducing the thrombus volume by 9.6%. Increasing the SG-FRT distance in patient 2 resulted in faster thoracic thrombosis and increased total thrombus volume.

Conclusion:

The location of reentry tears can influence the progression of FL thrombosis following TEVAR. The more distal the reentry tear in the aorta the more likely it is that FL thrombosis will occur. Hence, the distal landing zone of the stent-graft should be chosen carefully to ensure a sufficient SG-FRT distance.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Trombose / Aneurisma da Aorta Torácica / Implante de Prótese Vascular / Procedimentos Endovasculares / Hemodinâmica / Dissecção Aórtica Tipo de estudo: Etiology_studies / Prognostic_studies / Risk_factors_studies Limite: Humans Idioma: En Ano de publicação: 2020 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Trombose / Aneurisma da Aorta Torácica / Implante de Prótese Vascular / Procedimentos Endovasculares / Hemodinâmica / Dissecção Aórtica Tipo de estudo: Etiology_studies / Prognostic_studies / Risk_factors_studies Limite: Humans Idioma: En Ano de publicação: 2020 Tipo de documento: Article