Your browser doesn't support javascript.
loading
The Impact of Novel X-Ray Systems and X-Ray System Optimization on Patient Radiation Dose Administered During Cardiac Catheterization.
J Invasive Cardiol ; 32(6): 218-221, 2020 Jun.
Article em En | MEDLINE | ID: mdl-32385193
ABSTRACT

INTRODUCTION:

The effect of x-ray system optimization on patient radiation dose has received limited study.

METHODS:

We analyzed patient radiation dose in 1786 cardiac catheterization procedures (diagnostic coronary angiography and/or percutaneous coronary intervention [PCI]) performed at a single tertiary-care center before and after x-ray system optimization.

RESULTS:

After optimization, cineangiography dose-area product (DAP) dose was lower in the overall group of patients who underwent diagnostic angiography and/or PCI (1347 µGy•m² [IQR, 645-2345 µGy•m²] vs 1658 µGy•m² [IQR, 640- 2757 µGy•m²]; P=.03), as well as in the diagnostic angiography group (1795 µGy•m² [IQR, 1140-2994 µGy•m²] vs 2356 µGy•m² [IQR, 311-3576 µGy•m²]; P<.01) and PCI group (2152 µGy•m² [IQR, 1338-3477 µGy•m²] vs 2562 µGy•m² [IQR, 1681-3859 µGy•m²]; P=.02). Cineangiography DAP per exposure was also lower in the overall group (143 µGy•m² [IQR, 91-212 µGy•m²] vs 164 µGy•m² [IQR, 106-233 µGy•m²] per exposure; P<.01) and in the diagnostic angiography group (158 µGy•m² [IQR, 102-225 µGy•m²] vs 184 µGy•m² [IQR, 125-271 µGy•m²] per exposure; P<.01). After optimization, cineangiography air kerma (AK) dose (319 mGy [IQR, 197-531 mGy] vs 421 mGy [IQR, 241-600 mGy]; P=.01) and cineangiography AK per exposure (20.7 mGy [IQR, 12.9-29.0 mGy] vs 23.6 mGy [IQR, 14.1-32.9 mGy] per exposure; P=.03) were also lower in the PCI group. There was no significant change in fluoroscopy AK dose after optimization (20.7 mGy [IQR, 12.7-30.1 mGy] vs 20.4 mGy [IQR, 12.8-31.3 mGy] per minute; P=.71) and fluoroscopy DAP dose (156 µGy•m² [IQR, 101-242 µGy•m²] vs 156 µGy•m² [IQR, 102-236 µGy•m²] per minute; P=.91).

CONCLUSION:

X-ray system optimization was associated with lower cineangiography DAP, but similar fluoroscopy radiation dose.
Assuntos
Palavras-chave
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Exposição à Radiação / Intervenção Coronária Percutânea Limite: Humans Idioma: En Ano de publicação: 2020 Tipo de documento: Article
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Exposição à Radiação / Intervenção Coronária Percutânea Limite: Humans Idioma: En Ano de publicação: 2020 Tipo de documento: Article