A Multicenter Evaluation of a Nongel Mechanical Separator Plasma Blood Collection Tube for Testing of Selected Therapeutic Drugs.
Morosyuk, Svetlana; Berube, Julie; Christenson, Robert; Wu, Alan H B; Uettwiller-Geiger, Denise; Palicka, Vladimir; Prusa, Richard; Zima, Tomas; Shaw, Kristin; Plokhoy, Elizabeth; Ahuja, Aparna J; Stankovic, Ana K.
Afiliação
  • Morosyuk S; BD, Franklin Lakes, NJ.
  • Berube J; BD, Franklin Lakes, NJ.
  • Christenson R; School of Medicine, University of Maryland, MD, USA.
  • Wu AHB; University of California, San Francisco, San Francisco, CA.
  • Uettwiller-Geiger D; John T. Mather Memorial Hospital, Port Jefferson, NY.
  • Palicka V; University Hospital, Hradec Krάlόve, Czech Republic.
  • Prusa R; University Hospital Motol, Prague, Czech Republic.
  • Zima T; General University Hospital, Prague, Czech Republic.
  • Shaw K; School of Medicine, University of Maryland, MD, USA.
  • Plokhoy E; BD, Franklin Lakes, NJ.
  • Ahuja AJ; BD, Franklin Lakes, NJ.
  • Stankovic AK; BD, Franklin Lakes, NJ.
J Appl Lab Med ; 5(4): 671-685, 2020 07 01.
Article em En | MEDLINE | ID: mdl-32603441
ABSTRACT

BACKGROUND:

Some therapeutic drugs are unstable during sample storage in gel tubes. BD Vacutainer® Barricor™ Plasma Blood Collection Tube with nongel separator was compared with plasma gel tubes, BD Vacutainer PST™, PST II, and BD Vacutainer Serum Tube for acetaminophen, salicylate, digoxin, carbamazepine, phenytoin, valproic acid, and vancomycin during sample storage for up to 7 days.

METHODS:

Seven hospital sites enrolled 705 participants who were taking at least one selected drug. The study tubes were collected and tested at initial time (0 h), after 48 h of storage at room temperature and on day 7 (after additional 5 days of refrigerated storage). The performance of BD Barricor tube was evaluated for each drug by comparing BD Barricor samples with samples from the other tubes at 0 h from the same participant; stability was evaluated by comparing test results from the same tube at 0 h, 48 h, and 7 days.

RESULTS:

At 0 h, BD Barricor showed clinically equivalent results for selected therapeutic drugs compared with the other tubes, except phenytoin in BD PST. Phenytoin samples ≥20 µg/mL in BD PST had 10-12% lower values than samples in BD Barricor. During sample storage, all selected drugs remained stable for 7 days in BD Barricor and in serum aliquots. In BD PST, all drugs remained stable except phenytoin and carbamazepine and in BD PST II except for phenytoin.

CONCLUSION:

The BD Barricor Tube is effective for the collection and storage of plasma blood samples for therapeutic drug monitoring without sample aliquoting.