Characterization and Performance Evaluation of the First-Proton Therapy Facility in India.
Shamurailatpam, Dayananda Sharma; Manikandan, A; Ganapathy, K; Noufal, M P; Patro, Kartikeshwar C; Rajesh, T; Jalali, R.
Afiliação
  • Shamurailatpam DS; Department of Medical Physics, Apollo Proton Cancer Centre, Chennai, Tamil Nadu, India.
  • Manikandan A; Department of Medical Physics, Apollo Proton Cancer Centre, Chennai, Tamil Nadu, India.
  • Ganapathy K; Department of Medical Physics, Apollo Proton Cancer Centre, Chennai, Tamil Nadu, India.
  • Noufal MP; Department of Medical Physics, Apollo Proton Cancer Centre, Chennai, Tamil Nadu, India.
  • Patro KC; Department of Medical Physics, Apollo Proton Cancer Centre, Chennai, Tamil Nadu, India.
  • Rajesh T; Department of Medical Physics, Apollo Proton Cancer Centre, Chennai, Tamil Nadu, India.
  • Jalali R; Department of Radiation Oncology, Apollo Proton Cancer Centre, Chennai, Tamil Nadu, India.
J Med Phys ; 45(2): 59-65, 2020.
Article em En | MEDLINE | ID: mdl-32831487
ABSTRACT

PURPOSE:

The purpose of this study is to evaluate the performance characteristic of volumetric image-guided dedicated-nozzle pencil beam-scanning proton therapy (PT) system. MATERIALS AND

METHODS:

PT system was characterized for electromechanical, image quality, and registration accuracy. Proton beam of 70.2-226.2 MeV was characterized for short- and long-term reproducibility in integrated depth dose; spot profile characteristics at different air gap and gantry angle; positioning accuracy of single and pattern of spot; dose linearity, reproducibility and consistency. All measurements were carried out using various X-ray and proton-beam specific detectors following standard protocols.

RESULTS:

All electro-mechanical, imaging, and safety parameters performed well within the specified tolerance limit. The image registration errors along three translation and three rotational axes were ≤0.5 mm and ≤0.2° for both point-based and intensity-based auto-registration. Distal range (R90) and distal dose fall-off (DDF) of 70.2-226.2 MeV proton beams were within 1 mm of calculated values based on the international commission on radiation units and measurements 49 and 0.0156× R90, respectively. The R90 and DDF were reproducible within a standard deviation of 0.05 g/cm2 during the first 8 months. Dose were linear from 18.5 (0.011 MU/spot) to 8405 (5 MU/spot) MU, reproducible within 0.5% in 5 consecutive days and consistent within 0.8% for full rotation. The cGy/MU for 70.2-226.2MeV was consistent within 0.5%. In-air X(Y) spot-sigma at isocenter varies from 2.96 (3.00) mm to 6.68 (6.52) mm for 70.2-226.2 MeV. Maximum variation of spot-sigma with air-gap of ±20 cm was ±0.36 mm (5.28%) and ±0.82 mm (±12.5%) along X- and Y-direction and 3.56% for full rotation. Relative spot positions were accurate within ±0.6 mm. The planned and delivered spot pattern of known complex geometry agreed with (γ%≤1) for 1% @ 1 mm >98% for representative five-proton energies at four gantry angle.

CONCLUSION:

The PT-system performed well within the expected accuracy level and consistent over a period of 8 months. The methodology and data presented here may help upcoming modern PT center during their crucial phase of commissioning.