Your browser doesn't support javascript.
loading
Ready-to-use nanopore platform for the detection of any DNA/RNA oligo at attomole range using an Osmium tagged complementary probe.
Kang, Albert S W; Bernasconi, Janette G; Jack, William; Kanavarioti, Anastassia.
Afiliação
  • Kang ASW; Yenos Analytical LLC, 4659 Golden Foothill Pkwy, Suite 101, El Dorado Hills, CA, 95672, USA.
  • Bernasconi JG; Yenos Analytical LLC, 4659 Golden Foothill Pkwy, Suite 101, El Dorado Hills, CA, 95672, USA.
  • Jack W; New England BioLabs, Inc., Ipswich, MA, USA.
  • Kanavarioti A; Yenos Analytical LLC, 4659 Golden Foothill Pkwy, Suite 101, El Dorado Hills, CA, 95672, USA. tessi.kanavarioti@gmail.com.
Sci Rep ; 10(1): 19790, 2020 11 13.
Article em En | MEDLINE | ID: mdl-33188229
ABSTRACT
Nanopores can serve as single molecule sensors. We exploited the MinION, a portable nanopore device from Oxford Nanopore Technologies, and repurposed it to detect any DNA/RNA oligo (target) in a complex mixture by conducting voltage-driven ion-channel measurements. The detection and quantitation of the target is enabled by the use of a unique complementary probe. Using a validated labeling technology, probes are tagged with a bulky Osmium tag (Osmium tetroxide 2,2'-bipyridine), in a way that preserves strong hybridization between probe and target. Intact oligos traverse the MinION's nanopore relatively quickly compared to the device's acquisition rate, and exhibit count of events comparable to the baseline. Counts are reported by a publicly available software, OsBp_detect. Due to the presence of the bulky Osmium tag, probes traverse more slowly, produce multiple counts over the baseline, and are even detected at single digit attomole (amole) range. In the presence of the target the probe is "silenced". Silencing is attributed to a 11 double stranded (ds) complex that does not fit and cannot traverse this nanopore. This ready-to-use platform can be tailored as a diagnostic test to meet the requirements for point-of-care cell-free tumor DNA (ctDNA) and microRNA (miRNA) detection and quantitation in body fluids.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: DNA / Nanoporos Tipo de estudo: Diagnostic_studies Idioma: En Ano de publicação: 2020 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: DNA / Nanoporos Tipo de estudo: Diagnostic_studies Idioma: En Ano de publicação: 2020 Tipo de documento: Article