Your browser doesn't support javascript.
loading
An explicit expression for the retention factor and velocity dependency of the mobile zone mass transfer band broadening in packed spheres beds used in liquid chromatography.
Matheuse, Frederick; Deridder, Sander; Desmet, Gert.
Afiliação
  • Matheuse F; Department of Chemical Engineering, Vrije Universiteit Brussel, Brussels, Belgium.
  • Deridder S; Department of Chemical Engineering, Vrije Universiteit Brussel, Brussels, Belgium; Department of Chemical Engineering, Process and Environmental Technology Lab (PETLab), KU Leuven, Sint-Katelijne-Waver, Belgium.
  • Desmet G; Department of Chemical Engineering, Vrije Universiteit Brussel, Brussels, Belgium. Electronic address: gedesmet@vub.be.
J Chromatogr A ; 1634: 461710, 2020 Dec 20.
Article em En | MEDLINE | ID: mdl-33221656
The present study proposes a ready-to-use analytical expression to calculate the mobile zone mass transfer contribution (hCm) in packed bed columns. For this purpose, first high-accuracy computations of the band broadening in a perfectly ordered sphere array (fcc-arrangement, external porosity ε=0.40) were made using computational fluid dynamics (CFD), covering a broad range of zone retention factors (2≤k''≤18) and reduced velocities (0≤νi≤48). Subsequently, these data were used to determine the value of the geometrical constants in a number of possible analytical expressions for the hCm-contribution. This fitting exercise showed the traditional literature approach, using the Wilson-Geankoplis correlation to calculate the dimensionless Sherwood (Sh) number for the mass transfer, leads to fitting errors on the hCm-term as large as 150%. Instead, a new correlation for Sh is established. In addition, we also explored the difference in fitting accuracy between hCm-expressions based on either a plug-flow or a laminar flow profile assumption. Surprisingly, no significant difference in fitting accuracy between both assumptions was observed. Finally, a best-fit analytical expression is proposed that can represent the CFD-computed band broadening data with an average absolute fitting error of Δh=0.005, corresponding to a relative error of 2.5% on the hCm-term and of only 0.3% on the total plate height in a perfectly ordered sphere packing. Defining the presently investigated fcc-ordered sphere array with external porosity=40% as the reference geometry for a perfect sphere packing, the established expression can be used as a new yardstick expression against which the degree of eddy-dispersion can be measured.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Cromatografia Líquida / Hidrodinâmica Tipo de estudo: Prognostic_studies Idioma: En Ano de publicação: 2020 Tipo de documento: Article