Evaluation of a Motion Correction Algorithm for C-Arm Computed Tomography Acquired During Transarterial Chemoembolization.
Becker, Lena S; Gutberlet, Marcel; Maschke, Sabine K; Werncke, Thomas; Dewald, Cornelia L A; von Falck, Christian; Vogel, Arndt; Kloeckner, Roman; Meyer, Bernhard C; Wacker, Frank; Hinrichs, Jan B.
Afiliação
  • Becker LS; Department of Diagnostic and Interventional Radiology, Institute for Diagnostic and Interventional Radiology, Hannover Medical School, Carl-Neuberg-Str. 1, 30625, Hannover, Germany.
  • Gutberlet M; Department of Diagnostic and Interventional Radiology, Institute for Diagnostic and Interventional Radiology, Hannover Medical School, Carl-Neuberg-Str. 1, 30625, Hannover, Germany.
  • Maschke SK; Department of Diagnostic and Interventional Radiology, Institute for Diagnostic and Interventional Radiology, Hannover Medical School, Carl-Neuberg-Str. 1, 30625, Hannover, Germany.
  • Werncke T; Department of Diagnostic and Interventional Radiology, Institute for Diagnostic and Interventional Radiology, Hannover Medical School, Carl-Neuberg-Str. 1, 30625, Hannover, Germany.
  • Dewald CLA; Department of Diagnostic and Interventional Radiology, Institute for Diagnostic and Interventional Radiology, Hannover Medical School, Carl-Neuberg-Str. 1, 30625, Hannover, Germany.
  • von Falck C; Department of Diagnostic and Interventional Radiology, Institute for Diagnostic and Interventional Radiology, Hannover Medical School, Carl-Neuberg-Str. 1, 30625, Hannover, Germany.
  • Vogel A; Department of Gastroenterology and Hepatology, Hannover Medical School, Hannover, Germany.
  • Kloeckner R; Department of Diagnostic and Interventional Radiology, Johannes Gutenberg-University Medical Centre, Mainz, Germany.
  • Meyer BC; Department of Diagnostic and Interventional Radiology, Institute for Diagnostic and Interventional Radiology, Hannover Medical School, Carl-Neuberg-Str. 1, 30625, Hannover, Germany.
  • Wacker F; Department of Diagnostic and Interventional Radiology, Institute for Diagnostic and Interventional Radiology, Hannover Medical School, Carl-Neuberg-Str. 1, 30625, Hannover, Germany.
  • Hinrichs JB; Department of Diagnostic and Interventional Radiology, Institute for Diagnostic and Interventional Radiology, Hannover Medical School, Carl-Neuberg-Str. 1, 30625, Hannover, Germany. Hinrichs.jan@mh-hannover.de.
Cardiovasc Intervent Radiol ; 44(4): 610-618, 2021 Apr.
Article em En | MEDLINE | ID: mdl-33280058
ABSTRACT

PURPOSE:

The aim of this retrospective study was to evaluate the feasibility of a motion correction 3D reconstruction prototype technique for C-arm computed tomography (CACT). MATERIAL AND

METHODS:

We included 65 consecutive CACTs acquired during transarterial chemoembolization of 54 patients (47 m,7f; 67 ± 11.3 years). All original raw datasets (CACTOrg) underwent reconstruction with and without volume punching of high-contrast objects using a 3D image reconstruction software to compensate for motion (CACTMC_bone;CACTMC_no bone). Subsequently, the effect on image quality (IQ) was evaluated using objective (image sharpness metric) and subjective criteria. Subjective criteria were defined by vessel geometry, overall IQ, delineation of tumor feeders, the presence of foreign material-induced artifacts and need for additional imaging, assessed by two independent readers on a 3-(vessel geometry and overall IQ) or 2-point scale, respectively. Friedman rank-sum test and post hoc analysis in form of pairwise Wilcoxon signed-rank test were computed and inter-observer agreement analyzed using kappa test.

RESULTS:

Objective IQ as defined by an image sharpness metric, increased from 273.5 ± 28 (CACTOrg) to 328.5 ± 55.1 (CACTMC_bone) and 331 ± 57.8 (CACTMC_no bone; all p < 0.0001). These results could largely be confirmed by the subjective analysis, which demonstrated predominantly good and moderate inter-observer agreement, with best agreement for CACTMC_no bone in all categories (e.g., vessel geometry CACTOrg κ = 0.51, CACTMC_bone κ = 0.42, CACTMC_no bone κ = 0.69).

CONCLUSION:

The application of a motion correction algorithm was feasible for all data sets and led to an increase in both objective and subjective IQ parameters. LEVEL OF EVIDENCE 3.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Algoritmos / Tomografia Computadorizada por Raios X / Quimioembolização Terapêutica / Artefatos / Carcinoma Hepatocelular / Imageamento Tridimensional / Neoplasias Hepáticas Tipo de estudo: Diagnostic_studies / Observational_studies / Prognostic_studies Limite: Aged / Female / Humans / Male Idioma: En Ano de publicação: 2021 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Algoritmos / Tomografia Computadorizada por Raios X / Quimioembolização Terapêutica / Artefatos / Carcinoma Hepatocelular / Imageamento Tridimensional / Neoplasias Hepáticas Tipo de estudo: Diagnostic_studies / Observational_studies / Prognostic_studies Limite: Aged / Female / Humans / Male Idioma: En Ano de publicação: 2021 Tipo de documento: Article