Accuracy of a chairside intraoral scanner compared with a laboratory scanner for the completely edentulous maxilla: An in vitro 3-dimensional comparative analysis.
Zarone, Fernando; Ruggiero, Gennaro; Ferrari, Marco; Mangano, Francesco; Joda, Tim; Sorrentino, Roberto.
Afiliação
  • Zarone F; Professor and Chair, Division of Prosthodontics and Digital Dentistry, Department of Neurosciences, Reproductive and Odontostomatological Sciences, University "Federico II" of Naples, Naples, Italy.
  • Ruggiero G; PhD student, Division of Prosthodontics and Digital Dentistry, Department of Neurosciences, Reproductive and Odontostomatological Sciences, University "Federico II" of Naples, Naples, Italy. Electronic address: gennaro_ruggiero@hotmail.it.
  • Ferrari M; Professor and Chair, Division of Restorative Dentistry and Endodontics, Department of Medical Biotechnologies, University of Siena, Siena, Italy.
  • Mangano F; Associate Professor, Department of Prevention and Communal Dentistry, Sechenov First State Medical University of Moscow, Moscow, Russia.
  • Joda T; Professor and Vice Chair, Department of Reconstructive Dentistry, University Center for Dental Medicine Basel, University of Basel, Basel, Switzerland.
  • Sorrentino R; Research Professor, Division of Prosthodontics and Digital Dentistry, Department of Neurosciences, Reproductive and Odontostomatological Sciences, University "Federico II" of Naples, Naples, Italy.
J Prosthet Dent ; 124(6): 761.e1-761.e7, 2020 Dec.
Article em En | MEDLINE | ID: mdl-33289647
ABSTRACT
STATEMENT OF

PROBLEM:

Intraoral scanners are promising options for removable prosthodontics. However, analog aids, including occlusion rims, are still used, as a completely digital workflow is challenging and scientific evidence on the topic is scarce.

PURPOSE:

The purpose of this in vitro study was to assess and compare the trueness and precision of scans obtained from a reference typodont of a completely edentulous maxilla by using an intraoral scanner (TRIOS 3 Pod; 3Shape A/S) with scans obtained by using a laboratory scanner (DScan 3; EGS S.R.L.) from both Type IV stone casts and polysulfide impressions. MATERIAL AND

METHODS:

The polyurethane resin reference typodont was replicated from a clinical cast and was scanned with a metrological machine to obtain a reference scan. Ten digital casts were obtained by applying standardized scanning strategies to the reference typodont with the intraoral scanner. A device was created to make 10 consistent polysulfide impressions, and a scan of each impression was made with the laboratory scanner and then digitally reversed to obtain 10 digital reversed casts. Ten Type IV stone casts were poured and then scanned with the laboratory scanner to obtain 10 digital extraoral scanner casts. The scans in standard tessellation language (STL) format were imported into a dedicated software program, and the trueness and precision were calculated in µm. In addition to descriptive statistics (confidence interval 95%), 1-way ANOVA followed by the Bonferroni test or the Kruskal-Wallis and the Dunn tests were used to analyze differences among groups (α=.05).

RESULTS:

The trueness values (95% confidence interval) were digital intraoral scanner cast=48.7 (37.8-59.5), digital reversed cast=249.9 (121.3-378.5), and digital extraoral scanner cast=308.8 (186.6-430.9); significant differences were detected between digital intraoral scanner cast and digital reversed cast (P<.001) and between digital IOS casts and digital extraoral scanner cast (P<.001). The precision values (95% confidence interval) were digital intraoral scanner cast=46.7 (29.7-63.7), digital reversed cast=271.2 (94.6-447.8), and digital extraoral scanner cast=341.4 (175.5-507.3); significant differences were detected between digital intraoral scanner cast and digital reversed cast (P=.003) and between digital intraoral scanner cast and digital extraoral scanner cast (P=.001).

CONCLUSIONS:

Directly scanning a solid typodont of a completely edentulous maxilla with the intraoral scanner produced better trueness and precision than scanning the polysulfide impressions or the stone casts with a laboratory scanner.