Your browser doesn't support javascript.
loading
Use of compressed sensing to reduce scan time and breath-holding for cardiac cine balanced steady-state free precession magnetic resonance imaging in children and young adults.
Naresh, Nivedita K; Malone, LaDonna; Fujiwara, Takashi; Smith, Sarah; Lu, Quin; Twite, Mark D; DiMaria, Michael V; Fonseca, Brian M; Browne, Lorna P; Barker, Alex J.
Afiliação
  • Naresh NK; Section of Pediatric Radiology, Department of Radiology, Children's Hospital Colorado, University of Colorado Anschutz Medical Campus, 13123 E. 16th Ave., Aurora, CO, 80045, USA. nivedita.naresh@childrenscolorado.org.
  • Malone L; Section of Pediatric Radiology, Department of Radiology, Children's Hospital Colorado, University of Colorado Anschutz Medical Campus, 13123 E. 16th Ave., Aurora, CO, 80045, USA.
  • Fujiwara T; Section of Pediatric Radiology, Department of Radiology, Children's Hospital Colorado, University of Colorado Anschutz Medical Campus, 13123 E. 16th Ave., Aurora, CO, 80045, USA.
  • Smith S; Section of Pediatric Radiology, Department of Radiology, Children's Hospital Colorado, University of Colorado Anschutz Medical Campus, 13123 E. 16th Ave., Aurora, CO, 80045, USA.
  • Lu Q; Philips Medical Systems, Best, the Netherlands.
  • Twite MD; Section of Pediatric Anesthesiology, Department of Anesthesiology, Children's Hospital Colorado, University of Colorado Anschutz Medical Campus, Aurora, CO, USA.
  • DiMaria MV; Section of Pediatric Cardiology, Department of Pediatrics, Children's Hospital Colorado, University of Colorado Anschutz Medical Campus, Aurora, CO, USA.
  • Fonseca BM; Section of Pediatric Cardiology, Department of Pediatrics, Children's Hospital Colorado, University of Colorado Anschutz Medical Campus, Aurora, CO, USA.
  • Browne LP; Section of Pediatric Radiology, Department of Radiology, Children's Hospital Colorado, University of Colorado Anschutz Medical Campus, 13123 E. 16th Ave., Aurora, CO, 80045, USA.
  • Barker AJ; Section of Pediatric Radiology, Department of Radiology, Children's Hospital Colorado, University of Colorado Anschutz Medical Campus, 13123 E. 16th Ave., Aurora, CO, 80045, USA.
Pediatr Radiol ; 51(7): 1192-1201, 2021 Jun.
Article em En | MEDLINE | ID: mdl-33566124
ABSTRACT

BACKGROUND:

Conventional pediatric volumetric MRI acquisitions of a short-axis stack typically require multiple breath-holds under anesthesia.

OBJECTIVE:

Here, we aimed to validate a vendor-optimized compressed-sensing approach to reduce scan time during short-axis balanced steady-state free precession (bSSFP) cine imaging. MATERIALS AND

METHODS:

Imaging was performed in 28 patients (16±9 years) in this study on a commercial 3-tesla (T) scanner using retrospective electrocardiogram-gated cine bSSFP. Cine short-axis images covering both ventricles were acquired with conventional parallel imaging and a vendor-optimized parallel imaging/compressed-sensing approach. Qualitative Likert scoring for blood-myocardial contrast, edge definition, and presence of artifact was performed by two experienced radiologists. Quantitative comparisons were performed including biventricular size and function. A paired t-test was used to detect significant differences (P<0.05).

RESULTS:

Scan duration was 7±2 s/slice for conventional imaging (147±33 s total) vs. 4±2 s/slice for compressed sensing (83±28 s total). No significant differences were found with qualitative image scores for blood-myocardial contrast, edge definition, and presence of artifact. No significant differences were found in volumetric analysis between the two sequences. The number of breath-holds was 10±4 for conventional imaging and 5±3 for compressed sensing.

CONCLUSION:

Compressed sensing allowed for a 50% reduction in the number of breath-holds and a 43% reduction in the total scan time without differences in the qualitative or quantitative measurements as compared to the conventional technique.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Interpretação de Imagem Assistida por Computador / Imagem Cinética por Ressonância Magnética Tipo de estudo: Observational_studies / Qualitative_research Limite: Adult / Child / Humans Idioma: En Ano de publicação: 2021 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Interpretação de Imagem Assistida por Computador / Imagem Cinética por Ressonância Magnética Tipo de estudo: Observational_studies / Qualitative_research Limite: Adult / Child / Humans Idioma: En Ano de publicação: 2021 Tipo de documento: Article