Colon phantoms with cancer lesions for endoscopic characterization with optical coherence tomography.
Biomed Opt Express
; 12(2): 955-968, 2021 Feb 01.
Article
em En
| MEDLINE
| ID: mdl-33680552
Optical coherence tomography (OCT) is a growing imaging technique for real-time early diagnosis of digestive system diseases. As with other well-established medical imaging modalities, OCT requires validated imaging performance and standardized test methods for performance assessment. A major limitation in the development and testing of new imaging technologies is the lack of models for simultaneous clinical procedure emulation and characterization of healthy and diseased tissues. Currently, the former can be tested in large animal models and the latter can be tested in small animal disease models or excised human biopsy samples. In this study, a 23â
cm by 23â
cm optical phantom was developed to mimic the thickness and near-infrared optical properties of each anatomical layer of a human colon, as well as the surface topography of colorectal polyps and visual appearance compatible with white light endoscopy.
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Tipo de estudo:
Screening_studies
Idioma:
En
Ano de publicação:
2021
Tipo de documento:
Article