Berrington, Adam; Povazan, Michal; Mirfin, Christopher; Bawden, Stephen; Park, Young Woo; Marsh, Daniel C; Bowtell, Richard; Gowland, Penny A.
Afiliação
  • Berrington A; Sir Peter Mansfield Imaging Centre, School of Physics and Astronomy, University of Nottingham, Nottingham, UK.
  • Povazan M; Russell H. Morgan Department of Radiology, Johns Hopkins University School of Medicine, Baltimore, MD, USA.
  • Mirfin C; F. M. Kirby Research Center for Functional Brain Imaging, Kennedy Krieger Institute, Baltimore, MD, USA.
  • Bawden S; Sir Peter Mansfield Imaging Centre, School of Physics and Astronomy, University of Nottingham, Nottingham, UK.
  • Park YW; Sir Peter Mansfield Imaging Centre, School of Physics and Astronomy, University of Nottingham, Nottingham, UK.
  • Marsh DC; Center for Magnetic Resonance Research, Department of Radiology, University of Minnesota Medical School, Minneapolis, MN, USA.
  • Bowtell R; Sir Peter Mansfield Imaging Centre, School of Physics and Astronomy, University of Nottingham, Nottingham, UK.
  • Gowland PA; Sir Peter Mansfield Imaging Centre, School of Physics and Astronomy, University of Nottingham, Nottingham, UK.
Magn Reson Med ; 86(2): 611-624, 2021 08.
Article em En | MEDLINE | ID: mdl-33749010
ABSTRACT

PURPOSE:

Achieving a desired RF transmit field ( B1+ ) in small regions of interest is critical for single-voxel MRS at ultrahigh field. Radio-frequency (RF) shimming, using parallel transmission, requires B1+ mapping and optimization, which limits its ease of use. This work aimed to generate calibration-free RF shims for predefined target regions of interest, which can be applied to any participant, to produce a desired absolute magnitude B1+ (| B1+ |).

METHODS:

The RF shims were found offline by joint optimization on a database comprising B1+ maps from 11 subjects, considering regions of interest in occipital cortex, hippocampus and posterior cingulate, as well as whole brain. The | B1+ | achieved was compared with a tailored shimming approach, and MR spectra were acquired using tailored and calibration-free shims in 4 participants. Global and local 10g specific-absorption-rate deposition were estimated using Duke and Ella dielectric models.

RESULTS:

There was no difference in the mean | B1+ | produced using calibration-free versus tailored RF shimming in the occipital cortex (p = .15), hippocampus (p = .5), or posterior cingulate (p = .98), although differences were observed in the RMS error | B1+ |. Spectra acquired using calibration-free shims had similar SNR and low residual water signal. Under identical power settings, specific-absorption-rate deposition was lower compared with operating in quadrature mode. For example, the total head specific absorption rate was around 35% less for the occipital cortex.

CONCLUSION:

This work demonstrates that static RF shims, optimized offline for small regions, avoid the need for B1+ mapping and optimization for each region of interest and participant. Furthermore, power settings may be increased when using calibration-free shims, to better take advantage of RF shimming.