[Improved Design and Acceptance Test of Localization Couch for New Type of CT Simulator].
Zhongguo Yi Liao Qi Xie Za Zhi
; 45(2): 231-236, 2021 Apr 08.
Article
em Zh
| MEDLINE
| ID: mdl-33825389
Accurate CT simulation is the key link of precision radiotherapy, and the performance of the localization couch of CT simulator directly affects the accuracy of radiotherapy. With the rapid development of precision radiotherapy, conventional large aperture radiotherapy special CT simulator is difficult to meet the needs of precision radiotherapy localization, so most radiotherapy centers choose high-end diagnostic CT machines equipped with a flat tabletop for radiotherapy localization. In clinical work, the performance testing of the CT simulator localization couch is easy to be ignored. In addition, there are some problems such as insufficient precision in transforming the cradle-shaped couch top of diagnostic CT into a special flat couch top for radiotherapy. This paper provided an in-depth description of the improved design and performance test of the localization couch of the first special GE Revolution CT simulator for radiotherapy introduced by West China Hospital of Sichuan University. After the improvement, all the acceptance tests of the localization couch are in line with the standard, and the performance meets the high-precision radiotherapy localization needs of patients with different body weight in the center.
Palavras-chave
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Planejamento da Radioterapia Assistida por Computador
/
Tomografia Computadorizada por Raios X
Limite:
Humans
País/Região como assunto:
Asia
Idioma:
Zh
Ano de publicação:
2021
Tipo de documento:
Article