ABSTRACT
Aim:
Our objective was to test the feasibility of developing an LC-free, MS-based approach for high-throughput bioanalysis of humanized therapeutic monoclonal antibodies.Methodology:
A universal tryptic peptide from human IgG1, IgG3 and IgG4 was selected as the surrogate peptide for quantitation. After tryptic digestion, the surrogate peptide was fractionated via solid-phase extraction before being subjected to direct infusion-based MS/MS analysis. A high-resolution, multiplexed (MSX = 2) parallel reaction monitoring method was developed for data acquisition. Results &conclusion:
This proof-of-concept study demonstrated the feasibility of achieving high-throughput MS-based bioanalysis of monoclonal antibodies using an LC-free workflow with sensitivity comparable to conventional LC-MS/MS-based methods.Palavras-chave