Your browser doesn't support javascript.
loading
Long-Term Outcome of Small Head Metal-On-Metal Total Hip Arthroplasty: A 15-to-22 Year Follow-Up.
Maliarov, Anton; Huk, Olga L; Epure, Laura M; Bergeron, Stephane G; Antoniou, John; Zukor, David J.
Afiliação
  • Maliarov A; Department of Orthopedics, SMBD-Jewish General Hospital, Montreal, QC, Canada.
  • Huk OL; Department of Orthopedics, SMBD-Jewish General Hospital, Montreal, QC, Canada.
  • Epure LM; Department of Orthopedics, SMBD-Jewish General Hospital, Montreal, QC, Canada.
  • Bergeron SG; Department of Orthopedics, SMBD-Jewish General Hospital, Montreal, QC, Canada.
  • Antoniou J; Department of Orthopedics, SMBD-Jewish General Hospital, Montreal, QC, Canada.
  • Zukor DJ; Department of Orthopedics, SMBD-Jewish General Hospital, Montreal, QC, Canada.
J Arthroplasty ; 36(9): 3214-3220, 2021 Sep.
Article em En | MEDLINE | ID: mdl-33941413
ABSTRACT

BACKGROUND:

The Metasul articular interface was a second-generation metal-on-metal (MoM) total hip arthroplasty (THA) that was introduced as a promising interface with improved manufacturing technology, better clearances, and enhanced metal hardness. In December 2001, the manufacturer recalled these implants due to the failure of cup osseointegration.

METHODS:

Between 1997 and 2004, 168 consecutive primary Metasul THAs were performed in 144 patients. Most patients received a cementless femoral and porous-coated acetabular component with 28 mm head. A competing risk analysis was performed for determination separately for bearing surface-related, recalled bearing failure, and end point revision for any reason. For clinical patient evaluation, we used Harris hip score and University of California at Los Angeles scores. Cobalt and chromium ion level measurement and standard radiographic assessment was performed.

RESULTS:

Of the 168 THAs, 19 hips were revised at a mean period of 15 years as follows 12 of them were due to recalled acetabular component, five hips had noninterface-related complication, and two true interface surface failure. The survival distribution function of all hips with revision for any reason was 88.4%, for bearing interface relation 98.8%, and 92.6% for recalled cups. The mean Harris hip score and University of California at Los Angeles scores were 85 and 6, respectively, and the median Co and Cr blood levels were 1.0 and 0.91 µg/L.

CONCLUSION:

Excluding the recalled components, Metasul articular interface has performed extremely well at a minimum follow-up of 15 years in this relatively young population. There were two interface-related revisions in the entire cohort.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Artroplastia de Quadril / Próteses Articulares Metal-Metal / Prótese de Quadril Tipo de estudo: Observational_studies / Prognostic_studies Limite: Humans Idioma: En Ano de publicação: 2021 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Artroplastia de Quadril / Próteses Articulares Metal-Metal / Prótese de Quadril Tipo de estudo: Observational_studies / Prognostic_studies Limite: Humans Idioma: En Ano de publicação: 2021 Tipo de documento: Article