Comparing calculated and experimental activity and dose values obtained from image-based quantification of 90Y SPECT/CT Data.
Lotter, Konrad; Diemling, Markus; Sohlberg, Antti; Wiedner, Hannah; Haug, Alexander; Maringer, Franz Josef.
Afiliação
  • Lotter K; Technische Universität Wien - Technical University of Vienna, Karlsplatz 13, 1040 Wien, Austria. Electronic address: konrad.lotter@student.tuwien.ac.at.
  • Diemling M; HERMES Medical Solutions, Skeppsbron 44, 111 30 Stockholm, Sweden.
  • Sohlberg A; HERMES Medical Solutions, Skeppsbron 44, 111 30 Stockholm, Sweden.
  • Wiedner H; Bundesamt für Eich- und Vermessungswesen, Arltgasse 35, 1160 Wien, Austria.
  • Haug A; Division of Nuclear Medicine, Medical University of Vienna, Spitalgasse 23, 1090 Wien, Austria.
  • Maringer FJ; Technische Universität Wien - Technical University of Vienna, Karlsplatz 13, 1040 Wien, Austria; Bundesamt für Eich- und Vermessungswesen, Arltgasse 35, 1160 Wien, Austria.
Z Med Phys ; 31(4): 378-387, 2021 Nov.
Article em En | MEDLINE | ID: mdl-33966943
ABSTRACT

PURPOSE:

Selective internal radiation therapy (SIRT) is a treatment for various kinds of liver tumours by injecting 90Y bearing microspheres into the liver vessels. To perform meaningful post-treatment dosimetry, quantitative imaging is performed.

METHODS:

This work uses a Monte-Carlo based reconstruction software with scatter and attenuation correction and collimator modelling that allows the quantification of 90Y bremsstrahlung SPECT/CT data. A dataset comprising 17 patients and measurements on a Jaszczak phantom, a NEMA IEC Body phantom and an anthropomorphic liver phantom are analysed and activities and dose values are acquired. These measured values are compared with applied activities and pre-treatment calculations, allowing to assess the quality of the SPECT reconstruction. A detailed uncertainty budget is presented, including uncertainties of the dose calibrator, the count rate, non-included interactions and other factors.

RESULTS:

The applied method is validated by finding measurements repeatable within the given uncertainty, and it is shown the influence of various parameters on the reconstruction process is negligible. Furthermore, activities and doses measured in the phantoms show good agreement with calculated values, if they are corrected for partial volume effects.

CONCLUSIONS:

The strict observation of metrological requirements and the creation of an uncertainty budget increase the reliability and traceability of this novel approach to 90Y dosimetry. It gives an example of successful voxel-based dosimetry based on quantitative 90Y SPECT/CT image data.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Tomografia Computadorizada por Raios X / Tomografia Computadorizada de Emissão de Fóton Único Tipo de estudo: Health_economic_evaluation / Prognostic_studies Limite: Humans Idioma: En Ano de publicação: 2021 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Tomografia Computadorizada por Raios X / Tomografia Computadorizada de Emissão de Fóton Único Tipo de estudo: Health_economic_evaluation / Prognostic_studies Limite: Humans Idioma: En Ano de publicação: 2021 Tipo de documento: Article