The Accuracy of Continuous Glucose Monitoring in the Medical Intensive Care Unit.
J Diabetes Sci Technol
; 16(6): 1550-1554, 2022 Nov.
Article
em En
| MEDLINE
| ID: mdl-34218715
OBJECTIVE: To assess the accuracy of continuous glucose monitoring (CGM) in medical intensive care unit (MICU) patients. METHODS: A Medtronic Enlite® sensor accuracy was assessed versus capillary blood glucose (CBG) and plasma glucose (PG) using the mean absolute relative difference (MARD), surveillance error grid (SEG) analysis and modified Bland-Altman plots. RESULTS: Using CBG as a reference, MARD was 6.6%. Overall, 99.7% of the CGM readings were within the "no risk" zone. No significant differences in accuracy were seen within vasopressor subgroups. Using PG as the reference, MARD was 8.8%. The surveillance error grid analysis showed 95.2% of glucose readings were within the "no risk" zone. There were no device-related adverse events. CONCLUSION: The CGM sensor showed acceptable accuracy in MICU patients, regardless of vasopressor use.