Evaluation of four surface surrogates for modeling lung tumor positions over several fractions in radiotherapy.
Wikström, Kenneth A; Isacsson, Ulf M; Nilsson, Kristina M; Ahnesjö, Anders.
Afiliação
  • Wikström KA; Medical Physics, Uppsala University Hospital, Uppsala, Sweden.
  • Isacsson UM; Medical Radiation Sciences, Department of Immunology, Genetics and Pathology, Uppsala University, Uppsala, Sweden.
  • Nilsson KM; Medical Physics, Uppsala University Hospital, Uppsala, Sweden.
  • Ahnesjö A; Medical Radiation Sciences, Department of Immunology, Genetics and Pathology, Uppsala University, Uppsala, Sweden.
J Appl Clin Med Phys ; 22(9): 103-112, 2021 Sep.
Article em En | MEDLINE | ID: mdl-34258853
ABSTRACT
Patient breathing during lung cancer radiotherapy reduces the ability to keep a sharp dose gradient between tumor and normal tissues. To mitigate detrimental effects, accurate information about the tumor position is required. In this work, we evaluate the errors in modeled tumor positions over several fractions of a simple tumor motion model driven by a surface surrogate measure and its time derivative. The model is tested with respect to four different surface surrogates and a varying number of surrogate and image acquisitions used for model training. Fourteen patients were imaged 100 times with cine CT, at three sessions mimicking a planning session followed by two treatment fractions. Patient body contours were concurrently detected by a body surface laser scanning system BSLS from which four surface surrogates were extracted; thoracic point TP, abdominal point AP, the radial distance mean RDM, and a surface derived volume SDV. The motion model was trained on session 1 and evaluated on sessions 2 and 3 by comparing modeled tumor positions with measured positions from the cine images. The number of concurrent surrogate and image acquisitions used in the training set was varied, and its impact on the final result was evaluated. The use of AP as a surface surrogate yielded the smallest error in modeled tumor positions. The mean deviation between modeled and measured tumor positions was 1.9 mm. The corresponding deviations for using the other surrogates were 2.0 mm (RDM), 2.8 mm (SDV), and 3.0 mm (TP). To produce a motion model that accurately models the tumor position over several fractions requires at least 10 simultaneous surrogate and image acquisitions over 1-2 minutes.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Radiocirurgia / Neoplasias Pulmonares Limite: Humans Idioma: En Ano de publicação: 2021 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Radiocirurgia / Neoplasias Pulmonares Limite: Humans Idioma: En Ano de publicação: 2021 Tipo de documento: Article