Your browser doesn't support javascript.
loading
Objective measurement of lung volume recruitment therapy: laboratory and clinical validation.
Naughton, Phoebe E; Sheers, Nicole; Berlowitz, David J; Howard, Mark E; McKim, Douglas A; Katz, Sherri L.
Afiliação
  • Naughton PE; Institute for Breathing and Sleep, Heidelberg, Victoria, Australia.
  • Sheers N; Physiotherapy, Austin Health, Heidelberg, Victoria, Australia.
  • Berlowitz DJ; Institute for Breathing and Sleep, Heidelberg, Victoria, Australia.
  • Howard ME; Physiotherapy, The University of Melbourne, Melbourne, Victoria, Australia.
  • McKim DA; Institute for Breathing and Sleep, Heidelberg, Victoria, Australia david.berlowitz@austin.org.au.
  • Katz SL; Physiotherapy, Austin Health, Heidelberg, Victoria, Australia.
BMJ Open Respir Res ; 8(1)2021 07.
Article em En | MEDLINE | ID: mdl-34326156
Lung volume recruitment manoeuvres are often prescribed to maintain respiratory health in neuromuscular disease. Unfortunately, no current system accurately records delivered dose. This study determined the performance characteristics of a novel, objective, manual lung volume recruitment bag counter ('the counter') with bench and healthy volunteer testing, as well as in individuals with neuromuscular disease. We undertook (1) bench test determination of activation threshold, (2) bench and healthy volunteer fidelity testing during simulated patient interface leak and different pressure compressions and (3) comparisons with self-report in individuals with neuromuscular disease. The data are reported as summary statistics, compression counts, percentage of recorded versus delivered compressions and concordance (Cohen's kappa (K) and absolute agreement). RESULTS: Minimum counter activation pressure under conditions of zero leak was 1.9±0.4 cm H2O. No difference was observed between the number of repetitions delivered and recorded during high airway pressure condition. Interface leak approximating 25% resulted in underestimation of repetition counts, and once the leak was at 50% or beyond, the counter recorded no activity. Faster sampling frequency collected data with more fidelity. Counter data agreed with diary self-report during community trials (16 participants, 960 participant days, 77% agreement, Cohen's Κ=0.66 and p<0.001). Disagreement typically favoured more diary reported (18%) than counter (5%) sessions. CONCLUSIONS: The performance characteristics of a new lung volume recruitment counter have been established in both laboratory and community settings. Objective usage and dosage data should accelerate new knowledge development and better translation of lung volume recruitment therapy into policy and practice.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Respiração com Pressão Positiva / Laboratórios Tipo de estudo: Guideline Limite: Humans Idioma: En Ano de publicação: 2021 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Respiração com Pressão Positiva / Laboratórios Tipo de estudo: Guideline Limite: Humans Idioma: En Ano de publicação: 2021 Tipo de documento: Article