Percutaneous radiofrequency ablation near large vessels in beagle livers: the impact of time and distance on the ablation zone.
Jiang, Binbin; Zhao, Kun; Yan, Kun; Wang, Song; Meng, Yuanfeng; Liu, Baojiang; Wu, Hao; Wang, Haiyue.
Afiliação
  • Jiang B; Key Laboratory of Carcinogenesis and Translational Research (Ministry of Education/Beijing), Department of Ultrasound, Peking University Cancer Hospital & Institute, Beijing, China.
  • Zhao K; Key Laboratory of Carcinogenesis and Translational Research (Ministry of Education/Beijing), Department of Ultrasound, Peking University Cancer Hospital & Institute, Beijing, China.
  • Yan K; Key Laboratory of Carcinogenesis and Translational Research (Ministry of Education/Beijing), Department of Ultrasound, Peking University Cancer Hospital & Institute, Beijing, China.
  • Wang S; Key Laboratory of Carcinogenesis and Translational Research (Ministry of Education/Beijing), Department of Ultrasound, Peking University Cancer Hospital & Institute, Beijing, China.
  • Meng Y; Department of Ultrasound, Peking University First Hospital, Beijing, China.
  • Liu B; Key Laboratory of Carcinogenesis and Translational Research (Ministry of Education/Beijing), Department of Interventional Therapy, Peking University Cancer Hospital & Institute, Beijing, China.
  • Wu H; Key Laboratory of Carcinogenesis and Translational Research (Ministry of Education/Beijing), Department of Ultrasound, Peking University Cancer Hospital & Institute, Beijing, China.
  • Wang H; Key Laboratory of Carcinogenesis and Translational Research (Ministry of Education/Beijing), Department of Pathology, Peking University Cancer Hospital & Institute, Beijing, China.
Int J Hyperthermia ; 38(1): 1263-1270, 2021.
Article em En | MEDLINE | ID: mdl-34404325
ABSTRACT

PURPOSE:

To investigate the effects of ablation time and distance between the radiofrequency ablation (RFA) electrode tip and a large vessel on the ablation zone in beagle livers.

METHODS:

Sixty-one percutaneous RFA coagulation zones were created near large vessels in 10 beagle livers in vivo. The ablated lesions were divided into four groups based on ablation time and distance between the electrode tip and a large vessel (group A, 3 min 0.5 cm; group B, 3 min 0 cm; group C, 5 min 0.5 cm; group D, 5 min 0 cm). The ablated area, long-axis diameters, short-axis diameters, and vessel wall injury were examined.

RESULTS:

With a fixed ablation time, the ablation zone created with the electrode tip at 0.5 cm from the large vessel was significantly larger than at 0 cm (p < .05). At a fixed distance between the electrode tip and vessel, the ablation zone created for 5 min was significantly larger than for 3 min (p < .05). The frequency of vessel wall injury in the 0 cm groups was significantly higher than that in the 0.5 cm groups (37.5% vs. 6.9%; p = .003, odds ratio, 7.43). The ratio of width to depth (Dw/Dz) was larger in the 0.5 cm groups than in the 0 cm groups (p < .001).

CONCLUSION:

The ablation zone increased with longer ablation times and greater distances between the RFA tip and large vessels for perivascular lesions. The distance between the needle tip and blood vessels is an important factor that affects the overall ablation outcome.