Bioaffinity Screening with a Rapid and Sample-Efficient Autosampler for Native Electrospray Ionization Mass Spectrometry.
Kaeslin, Jérôme; Brunner, Cyrill; Ghiasikhou, Sahar; Schneider, Gisbert; Zenobi, Renato.
Afiliação
  • Kaeslin J; Department of Chemistry and Applied Biosciences, ETH Zürich, Vladimir-Prelog-Weg 3, CH-8093 Zurich, Switzerland.
  • Brunner C; Department of Chemistry and Applied Biosciences, ETH Zürich, Vladimir-Prelog-Weg 3, CH-8093 Zurich, Switzerland.
  • Ghiasikhou S; Department of Chemistry and Applied Biosciences, ETH Zürich, Vladimir-Prelog-Weg 3, CH-8093 Zurich, Switzerland.
  • Schneider G; Department of Chemistry and Applied Biosciences, ETH Zürich, Vladimir-Prelog-Weg 3, CH-8093 Zurich, Switzerland.
  • Zenobi R; Department of Chemistry and Applied Biosciences, ETH Zürich, Vladimir-Prelog-Weg 3, CH-8093 Zurich, Switzerland.
Anal Chem ; 93(39): 13342-13350, 2021 10 05.
Article em En | MEDLINE | ID: mdl-34546705
ABSTRACT
Fast and efficient handling of ligands and biological targets are required in bioaffinity screening based on native electrospray ionization mass spectrometry (ESI-MS). We use a prototype microfluidic autosampler, called the "gap sampler", to sequentially mix and electrospray individual small molecule ligands together with a target protein and compare the screening results with data from thermal shift assay and surface plasmon resonance. In a first round, all three techniques were used for a screening of 110 ligands against bovine carbonic anhydrase II, which resulted in five mutual hits and some false positives with ESI-MS presumably due to the high ligand concentration or interferences from dimethyl sulfoxide. In a second round, 33 compounds were screened in lower concentrations and in a less complex matrix, resulting in only true positives with ESI-MS. Within a cycle time of 30 s, dissociation constants were determined within an order of magnitude accuracy consuming only 5 pmol of ligand and less than 15 pmol of protein per screened compound. In a third round, dissociation constants of five compounds were accurately determined in a titration experiment. Thus, the gap sampler can rapidly and efficiently be used for high-throughput screening.