Intravenous Fluid Bag as a Substitute for Gel Standoff Pad in Musculoskeletal Point-of-care Ultrasound.
Mount, Cristin; Taylor, David; Skinner, Carl; Grogan, Scott.
Afiliação
  • Mount C; Department of Emergency Medicine, Madigan Army Medical Center, Tacoma, WA 98431, USA.
  • Taylor D; Department of Emergency Medicine, Madigan Army Medical Center, Tacoma, WA 98431, USA.
  • Skinner C; Department of Emergency Medicine, Madigan Army Medical Center, Tacoma, WA 98431, USA.
  • Grogan S; Department of Emergency Medicine, Madigan Army Medical Center, Tacoma, WA 98431, USA.
Mil Med ; 188(5-6): e949-e952, 2023 05 16.
Article em En | MEDLINE | ID: mdl-34676409
ABSTRACT

INTRODUCTION:

Point-of-care ultrasound (POCUS) is a tool undergoing expanding use in military medicine, including routine inpatient, outpatient, and operational environments. Specific musculoskeletal POCUS examinations require additional equipment in the form of a standoff assist device to maximize image acquisition. These devices may not be readily available to POCUS users in more austere or resource-constrained environments. We devised a study to determine if intravenous fluid bags of various volumes could be substituted for standard standoff devices in musculoskeletal POCUS. MATERIALS AND

METHODS:

Sequential images of a soft tissue foreign body model, an interphalangeal joint, and a chest wall were taken using a gel standoff pad or water bath and compared to images acquired using three different sizes of intravenous fluid bags after removing excess air from the bags. Images were de-identified and scored for quality using a visual analog scale. We used a two-factor analysis of variance without replication to analyze the differences in image quality between standoff devices, with a P-value less than.05 considered statistically significant.

RESULTS:

We performed 13 POCUS studies and had a group of eight POCUS-trained physicians and physician assistants score these studies, resulting in 104 total quality scores. There was no significant difference in image quality between standoff devices for chest (P-value.280) and hand (P-value 0.947) images. We found a significant difference in image quality between standoff devices for the soft tissue foreign body model (P-value 0.039), favoring larger intravenous fluid bag standoff over standard devices.

CONCLUSIONS:

In the absence of a standard commercial gel standoff device or water bath, intravenous fluid bags of 50, 100, and 250 mL can facilitate quality image acquisition for musculoskeletal POCUS.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Médicos / Sistemas Automatizados de Assistência Junto ao Leito Tipo de estudo: Prognostic_studies Limite: Humans Idioma: En Ano de publicação: 2023 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Médicos / Sistemas Automatizados de Assistência Junto ao Leito Tipo de estudo: Prognostic_studies Limite: Humans Idioma: En Ano de publicação: 2023 Tipo de documento: Article