Characterisation of a synthetic diamond detector for end-to-end dosimetry in stereotactic body radiotherapy and radiosurgery.
Shaw, Maddison; Lye, Jessica; Alves, Andrew; Keehan, Stephanie; Lehmann, Joerg; Hanlon, Maximilian; Kenny, John; Baines, John; Porumb, Claudiu; Geso, Moshi; Brown, Rhonda.
Afiliação
  • Shaw M; Australian Clinical Dosimetry Service, Australian Radiation Protection and Nuclear Safety Agency, Yallambie, VIC, Australia.
  • Lye J; School of Health and Biomedical Science, RMIT University, Melbourne, VIC, Australia.
  • Alves A; Australian Clinical Dosimetry Service, Australian Radiation Protection and Nuclear Safety Agency, Yallambie, VIC, Australia.
  • Keehan S; Olivia Newtown John Cancer Wellness & Research Centre, Heidelberg, VIC, Australia.
  • Lehmann J; Australian Clinical Dosimetry Service, Australian Radiation Protection and Nuclear Safety Agency, Yallambie, VIC, Australia.
  • Hanlon M; Australian Clinical Dosimetry Service, Australian Radiation Protection and Nuclear Safety Agency, Yallambie, VIC, Australia.
  • Kenny J; Department of Radiation Oncology, Calvary Mater Newcastle, Newcastle, Australia.
  • Baines J; School of Science, RMIT University, Melbourne, Australia.
  • Porumb C; School of Mathematical and Physical Sciences, University of Newcastle, Australia.
  • Geso M; Institute of Medical Physics, University of Sydney, Australia.
  • Brown R; Australian Clinical Dosimetry Service, Australian Radiation Protection and Nuclear Safety Agency, Yallambie, VIC, Australia.
Phys Imaging Radiat Oncol ; 20: 40-45, 2021 Oct.
Article em En | MEDLINE | ID: mdl-34722939
ABSTRACT
BACKGROUND AND

PURPOSE:

Synthetic diamond detectors offer real time measurement of dose in radiotherapy applications which require high spatial resolution. Additional considerations and corrections are required for measurements where the diamond detector is orientated at various angles to the incident beam. This study investigated diamond detectors for end-to-end testing of Stereotactic Body Radiotherapy (SBRT) and Stereotactic Radiosurgery (SRS) in the context of dosimetry audits. MATERIAL AND

METHODS:

Seven individual diamond detectors were investigated and compared with respect to warm up stability, dose-rate dependence, linearity, detector shadowing, energy response, cross-calibration, angular dependence and positional sensitivity in SBRT and SRS.

RESULTS:

Large variation in the cross calibration factors was found between the seven individual detectors. For each detector, the energy dependence in the cross calibration factor was on average <0.6% across the beam qualities investigated (Co-60 Gamma Knife, and MV beams with TPR20,10 0.684-0.733). The angular corrections for individual fields were up to 5%, and varied with field size. However, the average angular dependence for all fields in a typical SRS treatment delivery was <1%. The overall measurement uncertainty was 3.6% and 3.1% (2σ) for an SRS and SBRT treatment plan respectively.

CONCLUSION:

Synthetic diamond detectors were found to be reliable and robust for end-to-end dosimetry in SBRT and SRS applications. Orientation of the detector relative to the beam axis is an important consideration, as significant corrections are required for angular dependence.